Diagnostic Performance Evaluation of the Sd Bioline Malaria Antigen Ag Pf/pan Test (05fk60) in a Malaria Endemic Area of Southern Ethiopia
نویسندگان
چکیده
Rapid diagnostic tests (RDTs) capable of detecting and differentiating Plasmodium species are needed in areas in which microscopy is unsuitable. This study was conducted to assess the diagnostic performance of the rapid test kit - SD BIOLINE Malaria Ag Pf/Pan(r) (05FK60) in an endemic area. Microscopy of Giemsa-stained blood films were performed to detect and estimate the Plasmodium density in malaria suspected patients. The performance of the SD BIOLINE Malaria Ag Pf/Pan test was evaluated using 272 Plasmodium-positive and 102 negative blood samples. The overall sensitivity of the SD BIOLINE Malaria Ag Pf/Pan test was 99.5% for P. falciparum and 92.6% for non-P. falciparum malaria infections. The respective specificity, PPV, and NPV of the test were 98.0, 98.4, and 99.0% for the diagnosis of P. falciparum, and 100.0 %, 100.0%, and 94.4% for non-P. falciparum species. The SD BIOLINE Malaria Ag Pf/Pan test showed an excellent performance in diagnosing Plasmodium infections in an endemic setting. Therefore, this point-of-care test could be used as an alternative to microscopy in places where P. falciparum is endemic and microscopy is unsuitable.
منابع مشابه
Evaluation of two new immunochromatographic assays for diagnosis of malaria.
We assessed the performance of two new commercially available rapid diagnostic tests (RDTs) for malaria (SD Bioline Malaria Ag Pf test and Ag Pf/Pan test) in 200 patients with uncomplicated malaria between August and October 2007 in Madagascar. Results of the two RDTs were compared with those obtained by microscopy and real-time polymerase chain reaction. The sensitivity and specificity for det...
متن کاملPerformance of Paracheck™-Pf, SD Bioline malaria Ag-Pf and SD Bioline malaria Ag-Pf/pan for diagnosis of falciparum malaria in the Central African Republic
BACKGROUND Rapid diagnostic tests (RDTs) are the current complement to microscopy for ensuring prompt malaria treatment. We determined the performance of three candidate RDTs (Paracheck™-Pf, SD Bioline malaria Ag-Pf and SD Bioline malaria Ag-Pf/pan) for rapid diagnosis of malaria in the Central African Republic. METHODS Blood samples from consecutive febrile patients who attended for laborato...
متن کاملPerformance of SD Bioline Malaria Ag Pf/Pan rapid test in the diagnosis of malaria in South-Kivu, DR Congo
INTRODUCTION Use of malaria rapid diagnostic tests (RDTs) has improved the management of this disease. We evaluated the validity of the SD-Bioline Malaria-Ag-Pf/Pan™ (Batch 60952) RDT supplied by the Malaria Control Program of the DRCongo. METHODS cChildren (n = 460) aged below 5 years seen in curative care (CC) for suspected malaria and in pre-school consultation (PSC) in two rural centers u...
متن کاملField Performance of Wondfo and SD Bioline Malaria Pf/Pan Rapid Diagnostic Tests for Malaria Diagnosis in Koraput District, Odisha State, India
Rapid diagnostic tests (RDTs) have now been recommended and brought to field operation by the National Vector Borne Disease Control Programme (NVBDCP). The objective of this study was to evaluate the performance of two commonly used RDTs; Wondfo and SD Bioline Malaria Pf/Pan for malaria diagnosis in an endemic area of Koraput district, Odisha state. A total of 100 clinically suspected malaria p...
متن کاملEvaluation of Rapid Diagnostics for Plasmodium falciparum and P. vivax in Mae Sot Malaria Endemic Area, Thailand
Prompt and accurate diagnosis of malaria is the key to prevent disease morbidity and mortality. This study was carried out to evaluate diagnostic performance of 3 commercial rapid detection tests (RDTs), i.e., Malaria Antigen Pf/Pan™, Malaria Ag-Pf™, and Malaria Ag-Pv™ tests, in comparison with the microscopic and PCR methods. A total of 460 blood samples microscopically positive for Plasmodium...