Physics of rotating gamma systems for stereotactic radiosurgery.
نویسندگان
چکیده
PURPOSE The purpose of this study was to evaluate the characteristics of a new rotating gamma system for stereotactic radiosurgery by comparison with a well accepted system. METHODS AND MATERIALS A novel gamma unit for stereotactic radiosurgery has been developed and distributed to 15 hospitals in China. The unit contains 30 cobalt-60 gamma radiation sources with initial activity of 200 Ci (7.4 x 10(12) Bq) each. The sources are positioned along 30 arcs, and rotate continuously as a group in an axis orthogonal to the patient's body. Measurements have been made on a representative unit installed in the Auhai Radiosurgery Center at the Beijing Navy General Hospital in the People's Republic of China. Ionization chambers calibrated by an American accredited dosimetry calibration laboratory were used for these measurements, as well as radiochromic film and thermoluminescent dosimeters. The unit tested utilizes collimators of nominal diameters of 4, 8, 14, and 18 mm. Radiochromic film samples from a Leksell Model U Gamma Knife were evaluated by the same laboratory and are presented for comparison. The treatment planning system was not evaluated. RESULTS Radiation-absorbed dose rates and profiles measured for this unit are comparable to those previously measured with the same techniques for the Leksell Model U Gamma Knife units in San Diego and Atlanta. CONCLUSION This unit is capable of producing well collimated beams of high energy photons, suitable for stereotactic radiosurgery. It has similar physical characteristics to those previously reported for the Leksell Model U Gamma Knife unit.
منابع مشابه
Development of A Head Docking Device for Linac-based Stereotactic Radiosurgery System to be Installed on Neptun 10 PC Linac
Introduction: Stereotactic radiosurgery is a technique for the treatment of intracranial lesions with highly collimated x-ray beam. Linac-based radiosurgery is currently performed by two techniques: couch or pedestal mounted. In the first technique, a device is required to affix patient's head to a couch and moreover to translate it accurately. Structure of suc...
متن کاملDesign, Construction and Evaluation of an Anthropomorphic Head Phantom for Assessment of Image Distortion in Stereotactic Radiosurgery Planning Systems
Introduction: In recent years, the use of magnetic resonance (MR) images in radiation treatment planning has drawn considerable attention. However, although the extent of a tumor can be determined in great detail on MR images, the geometric accuracy of these images is limited by distortions stemming from the inhomogeneity of the static background magnetic field, the nonlineari...
متن کاملStereotactic Radiosurgery in the Treatment of Glioblastoma Multiforme: Current Status of Technology and Potential Role of Microbeam Radiosurgery
Stereotactic radiosurgery is an emerging treatment option offered to patients with glioblastoma (GB). Radiosurgery is performed as an outpatient procedure and provides a safe and effective non-invasive treatment for focal GB. High-energy beams originating from cobalt sources placed into a helmet-shaped device (Gamma Knife®, Elekta) or generated by a linear accelerator (LINAC), rotating on a gan...
متن کاملStereotactic Radiosurgery/Radiotherapy: A Historical Review
"Stereotactic" is an exact radiotherapy treatment modality which implements invasive and non-invasive facilities for improving precise dose delivery. Stereotactic refers to three-dimensional localization of a specific point in space by a unique set of coordinates that relate to a fixed external reference frame. An accurate delivery of radiation is attainable using these techniques with high pre...
متن کاملRotating and static sources for gamma knife radiosurgery systems: Monte Carlo studies.
Rotating gamma systems (RGSs), GammaART-6000, and its Chinese equivalents, such as OUR and MASEP, etc., are new radiosurgery systems that use rotating 60Co sources instead of the 201 static sources (Leksell gamma knife, LGK). The rotating sources of RGSs simulate an infinite number of beams and promote extremely high target to surface dose ratios. However, the results of Monte Carlo in this stu...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- International journal of radiation oncology, biology, physics
دوره 43 3 شماره
صفحات -
تاریخ انتشار 1999