A comparison of tracheal tube tip designs on the passage of an endotracheal tube during oral fiberoptic intubation.
نویسندگان
چکیده
BACKGROUND The design of an endotracheal tube has been shown to influence the passage of the tube through the glottis during fiberoptic intubation. Difficulty in passing the endotracheal tube can occur if the aryepiglottic folds obstruct the passage of the bevel. The relevant aspects of endotracheal tube design include the shape of the bevel, the material used by the manufacturer, and the ability of the tube to conform to the shape of the fiberscope. The aim of the current study was to compare the ease of passage through the glottis of two different tubes. One tube was a wire reinforced polyvinyl chloride tube with a standard bevel and the other was a newly designed tube with a bevel of different shape and made of silicone rubber. The new design is for use with the a commerical intubating laryngeal mask. METHODS The authors studied a population of 30 patients who received a standard anesthetic. In all cases, oral fiberoptic intubation was attempted. Anesthetic was administered to each patient using both tubes, and before the study the order of the tubes was randomized. The difficulty in passing the tube was assessed by a blinded observer and graded using a three-point scale (grade 1: no difficulty passing the tube; grade 2: obstruction to passing the tube relieved by withdrawal and a 90 degrees anticlockwise rotation; grade 3: obstruction necessitating more than one manipulation or external laryngeal manipulation). RESULTS In 27 patients, no difficulty was shown by use of the silicone-tipped tube. In only three patients was there difficulty that necessitated a 90 degrees anticlockwise twist. With the wire-reinforced tube, no difficulty was experienced on 14 occasions. Grade 1 difficulty was experienced eight times and difficulty necessitating more than one maneuver, head movement, or external laryngeal manipulation was seen on eight occasions. Statistical significance was achieved at P = 0.0002 (Wilcoxon signed rank test). CONCLUSIONS The authors conclude that the use of the silicone-tipped tube with the new bevel design may provide an advantage in the clinical situation of fiberoptic intubation.
منابع مشابه
Comparison of tracheal tube cuff inflation pressure among anesthesia team with different gradual level
Background & Objectives: Tracheal intubation is frequently performed in operating room. Over inflation of the endotracheal tube cuff is a risk factor for tracheal ischemia and subsequent complications. On the other hand inadequate cuff pressure increases the risk of aspiration of gastric contents. In this study, the endotracheal tube cuff pressure filled by the anesthesia team has been compared...
متن کاملComparison of two topographical airway length measurements in adults
BACKGROUND A correct estimate of the tracheal tube insertion depth can prevent complications, including endobronchial intubation and vocal cord trauma. We evaluated a new topographical method for endotracheal tube positioning relative to the carina, using a well-known prior topographical method for comparison. METHODS One hundred adult (male 50, female 50) patients were studied. The compariso...
متن کاملUnexpected difficulty in ventilating the lungs after tracheal intubation -A case report-
We experienced difficulty in ventilating the lungs of a patient after tracheal intubation. After intubation, an insufficient amount of tidal volume (V(T)) was delivered to the patient and the fiberoptic bronchoscopic examination identified partial abutment of the endotracheal tube (ETT) orifice against the tracheal wall. After various attempts to correctly place the ETT, a double-lumen endotrac...
متن کاملDifficult endotracheal intubation secondary to tracheal deviation and stenosis in a patient with severe kyphoscoliosis: a case report
We report on a case of difficult endotracheal intubation in a patient with marked tracheal deviation at an angle of 90 degrees combined with stenosis due to kyphoscoliosis with vertebral body fusion. After induction of general anesthesia, a proper laryngeal view was easily obtained using a videolaryngoscope. But a tracheal tube could not be advanced more than 3 cm beyond the vocal cords due to ...
متن کاملUnexpected tracheal web encountered during difficult intubation in the operating room.
A 60-year-old woman was prepared for revision of a previous anterior lumbar interbody fusion. After induction of anesthesia, three attempts to pass decreasing sizes of endotracheal tubes remained unsuccessful. An LMA Fastrach was inserted to maintain ventilation. Upon examination of the trachea with a flexible fiberoptic bronchoscope via the LMA Fastrach, a tracheal web was visualized 1 cm belo...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Anesthesiology
دوره 94 5 شماره
صفحات -
تاریخ انتشار 2001