3T Contrast-Enhanced Whole Heart Coronary MRA using 32-Channel Cardiac Coils for the Detection of Coronary Artery Disease

نویسندگان

  • Q. Yang
  • K. Li
  • X. Bi
  • J. An
  • F. Huang
  • R. Jerecic
  • D. Li
چکیده

Introduction: The development of contrast enhanced whole heart coronary MR angiography (CMRA) at 3T has shown promising results in clinical studies for the detection of significant coronary artery stenoses. However, the imaging time (~ 9 minutes) and spatial resolution (1.3 × 1.3 × 1.3 mm) remain major limitations. Newly available 32-channel cardiac coils allow greater acceleration factors, and thus reduced imaging time, reduced contrast agent dose and higher spatial resolution. Purpose: To evaluate the diagnostic accuracy of 3T contrast enhanced whole-heart CMRA with an optimized protocol for improved spatial resolution and reduced scan time using a 32-channel cardiac coils. Imaging time, image quality score, and diagnostic accuracy are evaluated in 32 consecutive patients with suspected coronary artery disease using conventional x-ray coronary angiography (CAG) as reference standard. Methods and Materials: 32 patients with suspected coronary artery disease who were scheduled for coronary angiography (CAG) (mean age 62 ± 12 y) underwent MRCA at 3T (MAGNETOM Tim Trio, Siemens) after informed consent was obtained. A 32-channel receiver coil was used for data acquisition (Invivo, Gainesville, FL). For image acquisition an ECG-triggered, navigator-gated, inversion-recovery prepared, segmented gradient-echo sequence was used with an acceleration factor of three in the phase-encoding direction using GRAPPA reconstruction. Imaging parameters included: voxel size 0.55x0.55x0.65 mm (interpolated from 1.1x1.1x1.3 mm), TR/TE = 3.3/1.5 msec, flip angle = 20°, bandwidth = 700 Hz/pixel, TI=200 msec. Contrast agent (0.15 mmol/kg body weight, MultiHance, Bracco, Italy) was intravenously administered at a rate of 0.3 ml/sec. For image analysis, standard 15-segment AHA classification system was used and only segments with a reference diameter of ≥ 1.5 mm were included, excluding those segments distal to complete occlusions. The diagnostic accuracy in detecting significant stenoses (≥50% of vessel lumen) was evaluated on the assessable segments only, as well as on all segments. Both non-assessable segments and the segments of the patients where the scan failed were considered to have a stenosis. Results: Whole-heart CMRA was successfully completed in 30 of 32 (94%) patients who were scheduled for CAG. The averaged imaging time was 5.9 ± 1.2 min. 40 of 392 segments (10%) with a reference luminal diameter > 1.5 mm on QCA were determined as non-assessable. On assessable segments only, the sensitivity, specificity, and accuracy of coronary MRA for detecting significant stenoses were 93% (78-99%), 99% (97-99%), 98%

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The clinical utility of contrast enhanced whole-heart coronary MRA with 32-channel coil at 3T scanner in the era of 64 and more-slice CT

Background Sixty-four and more-slice cardiac CT provides rapid and high quality coronary CT angiography for the majority of clinical needs of non-invasive evaluation of coronary artery. However, the recent advance of whole-heart coronary MRA has been able to provide high-quality of coronary MRA and also myocardial image in a reasonable scan time. The purpose of this study was to evaluate the cl...

متن کامل

3.0T whole-heart coronary magnetic resonance angiography performed with 32-channel cardiac coils: a single-center experience.

BACKGROUND Whole-heart coronary magnetic resonance angiography (MRA) is a promising method for noninvasive, radiation-free detection and exclusion of obstructive coronary artery disease; however, the required imaging time and robustness of the technique are not yet satisfactory. We evaluated the value of whole-heart coronary MRA at 3.0T using a 32-channel cardiac coil, which reduces image-acqui...

متن کامل

Diagnostic Accuracy of Different Image Postprocessing Methods for the Detection of Coronary Artery Stenoses by Using Contrast Enhanced Coronary MRA at 3.0T

Introduction: Contrast enhanced 3T whole heart coronary MR angiography (CMRA) is a reliable method for the detection and exclusion of hemodynamically significant coronary stenosis and has shown high accuracy in recent single center trial. Whole heart coronary MR angiography has been introduced as a method that can provide visualization of long segments of all coronary arteries with a single 3D ...

متن کامل

Diagnostic accuracy of 1.5-T unenhanced whole-heart coronary MR angiography performed with 32-channel cardiac coils: initial single-center experience.

PURPOSE To compare the imaging time and image quality obtained with whole-heart coronary magnetic resonance (MR) angiography performed with five- and 32-channel coils in healthy subjects and determine the accuracy of MR angiography performed with 32-channel coils in the detection of obstructive coronary artery disease (CAD). MATERIALS AND METHODS The institutional review board approved the st...

متن کامل

Diagnosis of Coronary Artery Disease via a Novel Fuzzy Expert System Optimized by Cuckoo Search

In this paper, we propose a novel fuzzy expert system for detection of Coronary Artery Disease, using cuckoo search algorithm. This system includes three phases: firstly, at the stage of fuzzy system design, a decision tree is used to extract if-then rules which provide the crisp rules required for Coronary Artery Disease detection. Secondly, the fuzzy system is formed by setting the intervals ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2008