Design of a new dual-helix loop coil for intravascular MR imaging
نویسندگان
چکیده
Introduction To discriminate between the vessel wall layers and plaque constituents in atherosclerosis, the image pixel size must be decreased to less than half a millimeter as the artery wall and fibrous cap thicknesses are very small (e.g. a fraction of a millimeter). For this purpose, researchers have introduced different types of intravascular loop and loopless coils, of which the single-helix loop coil is one. These coils are simply receiver antennas which have to be transported to the region of interest through a catheter and produce artery images with the highest pixel SNR in MRI. So far, some types of helix loop coils are presented as single-helix coils, single and array opposed-solenoid coils. In this study, a new dualhelix loop is introduced. Its SNR maps in radial and longitudinal planes have been analyzed and compared with a conventional single-helix loop coil. The results show that for the coil design introduced here, the amount of pixel SNR is increased by 33% compared to a conventional single-helix loop coil.
منابع مشابه
A New Intravascular Loopless Monopole Antenna (ILMA) for MR Imaging
In recent years, intravascular MR coil, which utilizes a special coil placed neighboring to the target artery, has been proposed to improve the local signal-to-noise ratio (SNR) and spatial resolution of vessel wall images. Especially the intravascular coil has identified its advantage in detecting atherosclerotic plaques [1]. At present, there are two types of intravascular MRI receivers: loop...
متن کاملA monopole/loop dual-tuned RF coil for ultrahigh field MRI.
Proton and heteronuclear MRI/MRS using dual-tuned (DT) coils could provide both anatomical and metabolic images without repositioning the subject. However, it is technologically challenging to attain sufficiently electromagnetic (EM) decoupling between the heteronuclear channel and proton channel, and keep the imaging areas and profiles of two nuclear channels highly matched. In this study, a h...
متن کاملRF-safe intravascular imaging using a self-visualizing transformer line
Objective: Intravascular MR-imaging may provide a local SNR gain possibly relevant for some interventional and diagnostic applications like high resolution intravascular imaging. However, regular signal transmission lines are not RF-safe due to possible RF heating of leads by resonant excitation of common mode currents. This safety issue prohibits intravascular MR imaging coils from being used ...
متن کاملConcentric dual-loop RF coil for magnetic resonance imaging∗
A surface coil for MRI consisted of two concentric loops was developed to produce brain images with a commercial MR imager. Prior to build the coil prototype, the magnetic field (B1) generated by the coil was numerically simulated. This field simulation is based on the Biot-Savart law for the circularand square-shaped loop coils. From these theoretical results, we can appreciate an improvement ...
متن کاملHigh-resolution intravascular magnetic resonance imaging: monitoring of plaque formation in heritable hyperlipidemic rabbits.
BACKGROUND The individual makeup of atherosclerotic plaque has been identified as a dominant prognostic factor. With the use of an intravascular magnetic resonance (MR) catheter coil, we evaluated the effectiveness of high-resolution MR in the study of the development of atherosclerotic lesions in heritable hyperlipidemic rabbits. METHODS AND RESULTS Sixteen hyperlipidemic rabbits were invest...