A simple technique to improve calculated skin dose accuracy in a commercial treatment planning system

نویسندگان

  • Lilie Wang
  • Anthony J. Cmelak
  • George X. Ding
چکیده

Radiation dermatitis during radiotherapy is correlated with skin dose and is a common clinical problem for head and neck and thoracic cancer patients. Therefore, accurate prediction of skin dose during treatment planning is clinically important. The objective of this study is to evaluate the accuracy of skin dose calculated by a commercial treatment planning system (TPS). We evaluated the accuracy of skin dose calculations by the anisotropic analytical algorithm (AAA) implemented in Varian Eclipse (V.11) system. Skin dose is calculated as mean dose to a contoured structure of 0.5 cm thickness from the surface. The EGSnrc Monte Carlo (MC) simulations are utilized for the evaluation. The 6, 10 and 15 MV photon beams investigated are from a Varian TrueBeam linear accelerator. The accuracy of the MC dose calculations was validated by phantom measurements with optically stimulated luminescence detectors. The calculation accuracy of patient skin doses is studied by using CT based radiotherapy treatment plans including 3D conformal, static gantry IMRT, and VMAT treatment techniques. Results show the Varian Eclipse system underestimates skin doses by up to 14% of prescription dose for the patients studied when external body contour starts at the patient's skin. The external body contour is used in a treatment planning system to calculate dose distributions. The calculation accuracy of skin dose with Eclipse can be considerably improved to within 4% of target dose by extending the external body contour by 1 to 2 cm from the patient's skin. Dose delivered to deeper target volumes or organs at risk are not affected. Although Eclipse treatment planning system has its limitations in predicting patient skin dose, this study shows the calculation accuracy can be considerably improved to an acceptable level by extending the external body contour without affecting the dose calculation accuracy to the treatment target and internal organs at risk. This is achieved by moving the calculation entry point away from the skin.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of Dose Calculation Accuracy of Isogray Treatment Planning System in Craniospinal Radiotherapy

Introduction: Craniospinal radiotherapy is a therapeutic technique for central nervous system (CNS) tumors, which requires meticulous attention to technique and dosimetry.Treatment planning system (TPS) is one of the main equipment in radiotherapy; therefore, the evaluation of its accuracy is essential for dose calculation. The present study evaluates the validity of Isogray TPS in craniospinal...

متن کامل

Assessment of out-of-field dose calculation algorithm by commercial treatment planning systems in IMRT and 3DCRT

Abstract Introduction: The accuracy an assessment of out- of- field dose due to secondary cancer risk is clinically important. Actually radiotherapy treatment planning systems are not commissioned for the out-of-field dose calculations, so the estimation of dose distributions by TPSs beyond the borders of treatment fields is not well calculate...

متن کامل

A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy

Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...

متن کامل

Dosimetric Evaluation of Dose calculation algorithms of Monaco Treatment Planning System in the heterogeneities area

Introduction: In radiation therapy, the accuracy of dose calculations by a treatment planning system (TPS) is important to achieve tumor control and to spare normal tissue. Treatment planning system calculations in the heterogeneous situation may present significant inaccuracies. In this study, three different dose calculation algorithms, pencil beam (PB), collapsed cone (CC), ...

متن کامل

Evaluation of Accuracy and Quality assurance of external beam therapy with photons

Introduction: Receiving exact dose by the patients is vital in radiotherapy. In radiation therapy, the dosimetry of radiations is too important because of successful radiation inquires for delivering the exact dose to the target volume. This study is to evaluate the tolerances and the accuracy of calculated dose of photon beams in the treatment software system. The TECDOC1583 p...