Interface dosimetry for electronic brachytherapy intracavitary breast balloon applicators

نویسندگان

  • James J. Segala
  • Gene A. Cardarelli
  • Jessica R. Hiatt
  • Bruce H. Curran
  • Edward S. Sternick
چکیده

In this study, we evaluate the attenuation of the dose due to barium-impregnation in the region between the surface of an electronic brachytherapy (EBT) balloon applicator for accelerated partial breast irradiation (APBI) and the prescription point at 1 cm depth in tissue. To perform the study, depth dose curves were calculated using a general purpose multi-particle transport code (FLUKA) for a range of balloon wall thicknesses with and without barium impregnation. Numerical data were verified with experimental readings using a parallel plate extrapolation ionization chamber for different wall thicknesses. Depth dose curves computed using both numerical and experimental methods show a 6.0% attenuation of the dose at the 1.0 cm prescription line due to the impregnation of barium in the balloon material, which agrees well with the manufacturer's specification. By applying this single attenuation factor, dose calculations throughout the entire planned volume are uniformly affected. However, at the balloon surface, attenuation on the order of 18.0% is observed. The AAPM TG-43 source data currently incorporated in commercially-available treatment planning systems do not account for the variable dose distributions attributable to balloon wall attenuation. Our results show that variable attenuation factors that may have clinical significance should be applied in order to determine near-surface dose distributions when using barium impregnated balloons for intracavitary breast brachytherapy. Dose distributions at distances greater than 1 cm from the surface of the balloon appear to be accurately represented without further modification.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Consideration of accurate dose distribution for Contura balloon catheter in breast brachytherapy with MCNP5 simulation

Introduction: Contura is a balloon applicator that used to deliver intra cavity radiation for Partial Breast Irradiation (PBI) treatments. Contura with 4 peripheral lumens surrounding the central, for loading HDR sources from an after loader system can provide an asymmetric dose distribution and gives a more flexible shape of dose than earlier balloon applicators such as Mammo...

متن کامل

Developing a Verification and Training Phantom for Gynecological Brachytherapy System

Introduction Dosimetric accuracy is a major issue in the quality assurance (QA) program for treatment planning systems (TPS). An important contribution to this process has been a proper dosimetry method to guarantee the accuracy of delivered dose to the tumor. In brachytherapy (BT) of gynecological (Gyn) cancer it is usual to insert a combination of tandem and ovoid applicators with a complicat...

متن کامل

Monte Carlo investigation on precise dosimetry of HDR breast brachytherapy with Accuboost

Introduction: Accuboost is a HDR brachytherapy system in early stages breast cancer treatment. This device provides a completely noninvasive procedure with parallel-opposed radiation from two immobilizing peripheral applicators that caused it a preferred option of modalities to choose. In most commercial treatment planning systems, tissues are considered as a simple water phan...

متن کامل

The use of tetrahedral mesh geometries in Monte Carlo simulation of applicator based brachytherapy dose distributions.

Accounting for brachytherapy applicator attenuation is part of the recommendations from the recent report of AAPM Task Group 186. To do so, model based dose calculation algorithms require accurate modelling of the applicator geometry. This can be non-trivial in the case of irregularly shaped applicators such as the Fletcher Williamson gynaecological applicator or balloon applicators with possib...

متن کامل

A Comparison of Skin Dose Delivered with MammoSite and Multicatheter Breast Brachytherapy

Background: Accelerated partial breast irradiation via interstitial balloon brachytherapy is a fast and effective treatment method for certain early stage breast cancers however skin, chest wall and Lung doses are correlated with toxicity in patients treated with breast brachytherapy.Objective: To investigate the percentage of the dose received by critical organ (skin), thermoluminescence detec...