Effects of manual hyperinflation, clinical practice versus expert recommendation, on displacement of mucus simulant: A laboratory study

نویسندگان

  • Marcia S Volpe
  • Juliane M Naves
  • Gabriel G Ribeiro
  • Gualberto Ruas
  • Mauro R Tucci
چکیده

INTRODUCTION Manual hyperinflation (MH), a maneuver applied in mechanically ventilated patients to facilitate secretion removal, has large variation in its performance. Effectiveness of MH is usually evaluated by its capacity to generate an expiratory flow bias. The aim of this study was to compare the effects of MH-and its resulting flow bias-applied according to clinical practice versus according to expert recommendation on mucus movement in a lung model simulating a mechanically ventilated patient. METHODS Twelve physiotherapists were asked to apply MH, using a self-inflating manual resuscitator, to a test lung as if to remove secretions under two conditions: according to their usual clinical practice (pre-instruction phase) and after verbal instruction to perform MH according to expert recommendation was given (post-instruction phase). Mucus simulant movement was measured with a photodensitometric technique. Peak inspiratory flow (PIF), peak inspiratory pressure (PIP), inspiratory time (TINSP), tidal volume (VT) and peak expiratory flow (PEF) were measured continuously. RESULTS It was found that MH performed post-instruction delivered a smaller VT (643.1 ± 57.8 ml) at a lower PIP (15.0 ± 1.5 cmH2O), lower PIF (38.0 ± 9.6 L/min), longer TINSP (1.84 ±0.54 s) and lower PEF (65.4 ± 6.7L/min) compared to MH pre-instruction. In the pre-instruction phase, MH resulted in a mean PIF/PEF ratio of 1.73 ± 0.38 and mean PEF-PIF difference of -54.6 ± 28.3 L/min, both out of the range for secretion removal. In the post-instruction phase both indexes were in the adequate range. Consequently, the mucus simulant was moved outward when MH was applied according to expert recommendation and towards the test lung when it was applied according to clinical practice. CONCLUSIONS Performance of MH during clinical practice with PIF higher than PEF was ineffective to clear secretion in a lung model simulating a mechanically ventilated patient. In order to remove secretion, MH should result in an adequate expiratory flow bias.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Effect of Inspiratory Time and Lung Compliance on Flow Bias Generated During Manual Hyperinflation: A Bench Study.

BACKGROUND Manual hyperinflation can be used to assist mucus clearance in intubated patients. The technique's effectiveness to move mucus is underpinned by its ability to generate flow bias in the direction of expiration, and this must exceed specific thresholds. It is unclear whether the inspiratory times commonly used by physiotherapists generate sufficient expiratory flow bias based on previ...

متن کامل

Effects of Manual Lung Hyperinflation Using Rebreathing Bags on Cardiorespiratory Parameters in Intubated Adults

Manual hyperinflation is a technique that is commonly used by physiotherapists to provide a greater than baseline tidal volume to the lungs of mechanically ventilated patients. The use of this technique is based more on clinical experience than on scientific evidence as the physiological effects of manual hyperinflation are uncertain. This review uses evidence-based principles to investigate th...

متن کامل

Effect of Lung Manual Hyperinflation (MHI) on Oxygenation of Patients Following Abdominal Surgery and T-Tube Support

Background: Postoperative pulmonary complications (PPC) are of the major reasons for death. Prolonged mechanical ventilation (PMV) and delayed extubation are leading to the incidence of more seriously complications. The effect of hyperinflation has not been investigated in control of these complications in patients who have been weaned from mechanical ventilation and are undergoing T-tube suppo...

متن کامل

Experimental study on the efficiency and safety of the manual hyperinflation maneuver as a secretion clearance technique *,**

OBJECTIVE To evaluate, in a lung model simulating a mechanically ventilated patient, the efficiency and safety of the manual hyperinflation (MH) maneuver as a means of removing pulmonary secretions. METHODS Eight respiratory therapists (RTs) were asked to use a self-inflating manual resuscitator on a lung model to perform MH as if to remove secretions, under two conditions: as routinely appli...

متن کامل

The use of a pressure manometer enhances student physiotherapists' performance during manual hyperinflation.

The purpose of this study was to determine the effect of using a pressure manometer on the delivery of target airway pressures during manual hyperinflation by student physiotherapists in the laboratory and clinical environments. Manual hyperinflations were delivered under control and feedback conditions where the feedback condition involved manual hyperinflation with a pressure manometer. Compa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 13  شماره 

صفحات  -

تاریخ انتشار 2018