An in vivo radiographic evaluation of the accuracy of Apex and iPex electronic Apex locators.
نویسندگان
چکیده
The aim of this study was to evaluate in vivo the clinical applicability of two electronic apex locators (EALs) - Apex (Septodont) and iPex (NSK) - in different groups of human teeth by using radiography. The working lengths (WLs) of 100 root canals were determined electronically. The EAL to be used first was chosen randomly and a K-file was inserted into the root canal until the EAL display indicated the location of the apical constriction (0 mm). The K-file was fixed to the tooth and a periapical radiograph was taken using a radiographic film holder. The K-file was removed and the WL was measured. The same procedure was repeated using the other EAL. Radiographs were examined with the aid of a light-box with lens of ×4 magnification by two blinded experienced endodontists. The distance between the file tip and the root apex was recorded as follows: (A) +1 to 0 mm, (B) -0.1 to 0.5 mm, (C) -0.6 to 1 mm, (D) -1.1 to 1.5 mm, and (E) -1.6 mm or greater. For statistical purposes, these scores were divided into 2 subgroups according to the radiographic apex: acceptable (B, C, and D) and non-acceptable (A and E). Statistically significant differences were not found between the results of Apex and iPex in terms of acceptable and non-acceptable measurements (p>0.05) or in terms of the distance recorded from file tip and the radiographic apex (p>0.05). Apex and iPex EALs provided reliable measurements for WL determination for endodontic therapy.
منابع مشابه
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination
In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...
متن کاملEvaluation of the accuracy of two apex locators in endodontic treatment and retreatments: an ex vivo study
Introduction: Working length determination is important in successful endodontic treatment and retreatment.This study evaluated the accuracy of two electronic apex locators Root ZX and Raypex®6 (EALs) in determining the electronic working length (EWL) of the root canals in endodontic treatment and retreatment. Materials and Methods: Access cavities were prepared on forty extracted, singl...
متن کاملThe effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants
BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...
متن کاملارزیابی دقت دو دستگاه آپکس یاب الکترونیکی (Electronic Apex Locator) از نسل چهارم در تعیین طول کانال (In-vivo)
Background and Aim: Determining correct working length in root canal therapy is of great importance in successful treatment. The purpose of this study was to compare the accuracy of two electronic apex locators of fourth generation, Raypex4 and Apit7, in determining working length in vivo. Materials and Methods: In this test evaluation study performed at Dental School of Tehran University, a to...
متن کاملمقایسه دقت دستگاههای اپکس یاب الکترونیکی ZX Root،Novapex در مقایسه با رادیوگرافی در شرایط in vivo
Background and Aims: Working length determination and remaining this length is of great importance in root canal therapy. Recently, electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the sensitivity of Root ZX and Novapex electronic apex locators with radiography in a clinical study. Materials and...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Brazilian dental journal
دوره 23 1 شماره
صفحات -
تاریخ انتشار 2012