Metal-on-metal total hip arthroplasty: the concerns.
نویسنده
چکیده
The metal-on-metal bearing couple is having a resurgence in clinical applications seen in total hip and hip resurfacing technologies. The most noteworthy advantage of a metal-on-metal implant is the improved wear characteristics seen in vitro on wear simulators and in vivo with retrieved implants. All bearings have disadvantages, and a metal-on-metal bearing is no exception. Concerns exist regarding the generation of metal ions seen in the blood and urine of patients with metal-on-metal implants. These elevated metal ions have theoretical, although not proven, risks related to carcinogenic and biologic concerns. Additionally, concerns exist regarding hypersensitivity, increased incidence of instability and increased costs. Specific patient selection issues arise with metal-on-metal implants. The current generation of implants has only early and mid-term results available, with no long-term series yet published. Therefore, although a metal-on-metal bearing may be considered a viable alternative to either polyethylene or ceramic implants, outstanding and unresolved issues continue to exist with this bearing, as they do with the alternatives.
منابع مشابه
A benign psoas mass following metal-on-metal resurfacing of the hip.
As metal-on-metal arthroplasty becomes more widespread, concerns are being raised about the potential dangers of metal particulate debris. We present the case of a benign psoas mass secondary to the presence of such particles. The mass was excised and the hip resurfacing subsequently revised to a total hip replacement.
متن کاملFatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature
Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...
متن کامل"Metal-on-metal total hip replacement. What does the literature say?".
Second-generation metal-on-metal (M/M) total hip replacements were introduced into clinical use in the late 1980s and demonstrate equivalent survivorship to conventional metal-on-polyethylene prostheses. Wear rates are comparable to those of first-generation designs that survived for a long time in the body. Biological effects from metal ions remain a concern. Patients with both first- and seco...
متن کاملMetal ion levels decrease after revision for metallosis arising from large-diameter metal-on-metal hip arthroplasty.
Concerns have been renewed regarding the possible long-term effects of elevated circulating levels of cobalt and chromium as a direct result of implantation of large femoral head diameter metal-on-metal bearings. In order to establish whether metal ion levels remain persistently elevated, we compared metal ion levels before and after revision surgery in patients with large head diameter (greate...
متن کاملDisappointing short-term results with the DePuy ASR XL metal-on-metal total hip arthroplasty.
Outcomes of ultralarge-diameter femoral heads used in metal-on-metal (MOM) total hip arthroplasty (THA) are relatively unknown. This study reports on early failures of the ASR XL (Depuy, Warsaw, Ind) and assesses whether a correlation with cup positioning exists. A retrospective review of 70 consecutive MOM THAs with ultralarge-diameter femoral head and monoblock acetabular component was conduc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical orthopaedics and related research
دوره 429 شماره
صفحات -
تاریخ انتشار 2004