A Comparison of Perimetric Results from a Tablet Perimeter and Humphrey Field Analyzer in Glaucoma Patients
نویسندگان
چکیده
PURPOSE To determine the correlation between the perimetric outcomes from perimetry software Melbourne Rapid Fields (MRF) run on an Apple iPad tablet and those from the Humphrey Field Analyzer (HFA). METHODS The MRF software was designed with features including variable fixation and fast thresholding using Bayes logic. Here, we report a cross-sectional study on 90 eyes from 90 participants: 12 had normal optic nerves and 78 had glaucoma with various degrees of visual field loss (41 mild and 37 moderate-severe). Exclusion criteria were patients with worse than 20/40 vision or recent intraocular surgery. The visual field outcomes of MRF were compared against those returned from the HFA 24-2 SITA standard. Participants were tested twice on the MRF to establish test-retest repeatability. RESULTS The test durations were shorter on MRF than HFA (5.7 ± 0.1 vs. 6.3 ± 0.1 minutes, P < 0.001). MRF showed a high level of concordance in its outcomes with HFA (intraclass coefficient [ICC] = 0.93 for mean defect [MD] and 0.86 for pattern deviation [PD]) although the MRF tended to give a less negative MD (1.4 dB bias) compared with the HFA. MRF also showed levels of test-retest reliability comparable to HFA (ICC = 0.93 for MD and 0.89 for PD, 95% limits of agreement -4.5 to 4.3 dB). CONCLUSION The perimetry results from the MRF have a strong correlation to the HFA outcomes. MRF also has test-retest reliability comparable to HFA. TRANSLATIONAL RELEVANCE Portable tablet perimetry may allow accurate assessment of visual field when standard perimetry machines are unavailable or unsuitable.
منابع مشابه
Investigating the effects of glaucomatous (POAG) damage on the mVEP parameters
Introduction: Glaucoma is considered as a major cause of irreversible vision loss, worldwide. Glaucoma includes a diverse ophthalmopathies characterized by attenuating the neural and connective tissue segments and eventually progression of specific patterns of visual dysfunction. Currently, perimetry is known as the most accurate diagnostic method in glaucoma and its follow up...
متن کاملA comparison of perimetric results with the Medmont and Humphrey perimeters.
BACKGROUND The Humphrey field analyser (HFA), Humphrey-Zeiss frequency doubling perimeter, and the Medmont automated perimeter (MAP) are three commonly used automated perimeters with threshold achromatic methodologies. Visual field loss may be detected earlier with strategies that target cell lines with reduced redundancy or which suffer selective damage. METHOD To compare these three perimet...
متن کاملSCIENTIFIC REPORT A comparison of global indices between the Medmont Automated Perimeter and the Humphrey Field Analyzer
Background: Two commonly used perimeters in Australia are the Humphrey Field Analyzer II (HFA) and the Medmont Automated Perimeter (MAP). Each device describes the visual field in terms of numerical values called global indices; however, these values are not interchangeable between devices. This study was designed to directly compare the global indices of HFA and MAP visual fields. Methods: 63 ...
متن کاملNew Technologies in Diagnosis of Primary Open-angle Glaucoma
The purpose of the study is to determine the most important diagnostic criteria for distinguishing preperimetric glaucoma from perimetric glaucoma on the basis of a comprehensive assessment of structural changes and indicators of ocular blood flow. 90 eyes, including 30 eyes with preperimetric glaucoma, 30 eyes with perimetric glaucoma, and 30 normal eyes were analyzed in the present study. The...
متن کامل[Functional relationship between retinal sensitivity threshold values assessed by standard automated perimetry in glaucoma].
OBJECTIVE To study the correlation between the sensitivity threshold values of the different points assessed by the Humphrey visual field analyzer (24-2 Swedish interactive threshold algorithm [SITA] standard strategy) in glaucoma patients. SUBJECTS, MATERIAL AND METHODS Prospective cross-sectorial study. One-hundred and four eyes of 104 glaucoma patients, defined by the appearance of the opt...
متن کامل