Dosimetric characterization of GMS BT‐125‐1 125I radioactive seed with Monte Carlo simulations and experimental measurement

نویسندگان

  • Nan Zhao
  • Ruijie Yang
  • Li Ren
  • Yi Fan
  • Junli Li
  • Jianguo Zhang
چکیده

PURPOSE To investigate the dosimetric characteristics of the new GMS BT-125-1 125 I radioactive seed, including dose rate constant, radial dose functions, and anisotropy functions. METHODS Dosimetric parameters of GMS BT-125-1 125 I seed including dose rate constant, radial dose functions, and anisotropy functions were calculated using the Monte Carlo code of MCNP5, and measured with thermoluminescent dosimeters (TLDs). The results were compared with those of PharmaSeed BT-125-1, PharmaSeed BT-125-2 125 I, and model 6711 125 I seeds. RESULTS The dose rate constant of GMS BT-125-1 125 I seed was 0.959 cGy·h-1·U-1, with the difference of 0.94%, 0.83%, and 0.73% compared with the PharmaSeed BT-125-1 125 I seed, PharmaSeed BT-125-2 125 I seed, and Model 6711 125 I seed, respectively. For radial dose function, the differences between the Monte Carlo and the experimental g(r) results were mostly within 10%. Monte Carlo results of g(r) for GMS BT-125-1 125 I seed were found in agreement (within 3.3%) with corresponding results for the PharmaSeed BT-125-2 125 I seed. The largest differences were 8.1% and 6.2% compared with PharmaSeed BT-125-1 125 I seed and model 6711 125 I seed, respectively. For anisotropy function, the difference between GMS BT-125-1 125 I seed and PharmaSeed BT-125-2 125 I seed was typically <10%. CONCLUSIONS The measured dose rate constant, radial dose functions, and two-dimensional anisotropy functions for the GMS BT-125-1 125 I seed showed good agreement with the Monte Carlo results. The dose rate constant of the GMS BT-125-1 125 I seed is similar to that of the PharmaSeed BT-125-1 125 I seed, the PharmaSeed BT-125-2 125 I seed, and the model 6711 125 I seed. For radial dose functions and two-dimensional anisotropy functions, the GMS BT-125-1 125 I seed is similar to the PharmaSeed BT-125-2 125 I seed but different from the PharmaSeed BT-125-1 125 I seed and the model 6711 125 I seed.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

تعیین پارامترهای دزیمتری یک چشمۀ طراحی‌شدۀ جدید 125I با مارکر Ag + Al2O3 در فانتوم آب

The purpose of this study is to determine the dosimetric parameters of a new 125I brachytherapy seed model, using Monte Carlo (MC) simulations according to the TG-43U1 recommendation. The Monte Carlo simulated dosimetric parameters of this sourse have been provided with MCNPX v.2.6 code. The source contains a cylindrical Ag+Al2O3 marker which is located at the center of the seed and onto which...

متن کامل

Measured transverse‐axis dosimetric parameters of the model STM1251 I125 interstitial source

The recent popularity of permanent implants for treatment of prostate cancer has created a large demand for low energy seed sources. Vendors have introduced new source designs to meet this demand. The AAPM has recommended that all low energy interstitial brachytherapy seed sources be subjected to independent dosimetric evaluations, preferably using experimental measurements as well as Monte Car...

متن کامل

Dosimetric parameters of three new solid core I‐125 brachytherapy sources

Monte Carlo calculations and TLD measurements have been performed for the purpose of characterizing dosimetric properties of new commercially available brachytherapy sources. All sources tested consisted of a solid core, upon which a thin layer of 125I has been adsorbed, encased within a titanium housing. The PharmaSeed BT-125 source manufactured by Syncor is available in silver or palladium co...

متن کامل

In vitro Dosimetric Study of Biliary Stent Loaded with Radioactive 125I Seeds

BACKGROUND A novel radioactive 125I seed-loaded biliary stent has been used for patients with malignant biliary obstruction. However, the dosimetric characteristics of the stents remain unclear. Therefore, we aimed to describe the dosimetry of the stents of different lengths - with different number as well as activities of 125I seeds. METHODS The radiation dosimetry of three representative ra...

متن کامل

Dosimetric comparison between four kinds of radioactive esophageal stents to be used in the treatment of advanced esophageal cancers using Monte Carlo simulation

Finding accurate methods to be employed for the treatment of esophageal cancers is of especial interest for researchers, due to the sensitivity of this tissue. Recently radioactive stents loaded with I-125 brachytherapy seeds have been widely investigated for the treatment of advanced esophageal cancer. It is necessary to investigate the dose distribution of any radioactive esophageal stents be...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 18  شماره 

صفحات  -

تاریخ انتشار 2017