Comparison of the Airtraq® and Truview® laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation in manikins
نویسندگان
چکیده
BACKGROUND Paramedics are frequently required to perform tracheal intubation, a potentially life-saving manoeuvre in severely ill patients, in the prehospital setting. However, direct laryngoscopy is often more difficult in this environment, and failed tracheal intubation constitutes an important cause of morbidity. Novel indirect laryngoscopes, such as the Airtraq and Truview laryngoscopes may reduce this risk. METHODS We compared the efficacy of these devices to the Macintosh laryngoscope when used by 21 Paramedics proficient in direct laryngoscopy, in a randomized, controlled, manikin study. Following brief didactic instruction with the Airtraq and Truview laryngoscopes, each participant took turns performing laryngoscopy and intubation with each device, in an easy intubation scenario and following placement of a hard cervical collar, in a SimMan manikin. RESULTS The Airtraq reduced the number of optimization manoeuvres and reduced the potential for dental trauma when compared to the Macintosh, in both the normal and simulated difficult intubation scenarios. In contrast, the Truview increased the duration of intubation attempts, and required a greater number of optimization manoeuvres, compared to both the Macintosh and Airtraq devices. CONCLUSION The Airtraq laryngoscope performed more favourably than the Macintosh and Truview devices when used by Paramedics in this manikin study. Further studies are required to extend these findings to the clinical setting.
منابع مشابه
A randomised crossover trial comparing the Airtraq(®) NT, McGrath(®) MAC and Macintosh laryngoscopes for nasotracheal intubation of simulated easy and difficult airways in a manikin.
OBJECTIVE Several devices can aid nasotracheal intubation when managing difficult airways. The McGrath MAC and Airtraq NT were compared with a Macintosh laryngoscope when studying the performance of anaesthetists with different levels of experience, in a manikin model of easy or difficult airway scenarios. METHODS Sixty-three anaesthetists were recruited into a randomised trial in which each ...
متن کاملComparison of the Glidescope® and Pentax AWS® laryngoscopes to the Macintosh laryngoscope for use by Advanced Paramedics in easy and simulated difficult intubation
BACKGROUND Intubation of the trachea in the pre-hospital setting may be lifesaving in severely ill and injured patients. However, tracheal intubation is frequently difficult to perform in this challenging environment, is associated with a lower success rate, and failed tracheal intubation constitutes an important cause of morbidity. Novel indirect laryngoscopes, such as the Glidescope and the A...
متن کاملEvaluation of intubation using the Airtraq or Macintosh laryngoscope by anaesthetists in easy and simulated difficult laryngoscopy--a manikin study.
The Airtraq Laryngoscope is a novel intubation device which allows visualisation of the vocal cords without alignment of the oral, pharyngeal and tracheal axes. We compared the Airtraq with the Macintosh laryngoscope in simulated easy and difficult laryngoscopy. Twenty-five anaesthetists were allowed up to three attempts to intubate the trachea in each of three laryngoscopy scenarios using a La...
متن کاملLearning curves of the Airtraq and the Macintosh laryngoscopes for tracheal intubation by novice laryngoscopists: a clinical study.
BACKGROUND The curved laryngoscope blade described by Macintosh in 1943 remains the most widely used device to facilitate tracheal intubation. The Airtraq laryngoscope is a new, single-use device for tracheal intubation. Several studies compared the use of Airtraq and Macintosh laryngoscopes in simulated intubation scenarios on manikins. We evaluated learning and performance of tracheal intubat...
متن کاملAirtraq vs standard laryngoscopy by student paramedics and experienced prehospital laryngoscopists managing a model of difficult intubation.
Two consecutive, randomised, cross-over trials compared intubation success rates in third-year paramedic students and experienced prehospital practitioners using the Airtraq or a Macintosh laryngoscope with flexible stylet in a manikin model of a Cormack and Lehane grade III/IV laryngoscopic view. First-time intubation rates for the Macintosh and Airtraq for students were 0/23 (0%) vs 10/23 (44...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- BMC Emergency Medicine
دوره 9 شماره
صفحات -
تاریخ انتشار 2009