Multi-site analytical evaluation of the Abbott ARCHITECT tacrolimus assay.
نویسندگان
چکیده
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Tacrolimus immunoassay. Proficiency panels and specimens from a population of organ transplant recipients were analyzed in 6 clinical laboratories in Europe and the United States, and the results were compared with other methods. The ARCHITECT assay requires a whole blood specimen pretreatment step with methanol/zinc sulfate to precipitate protein and extract the drug, followed by a 30-minute immunoassay using anti-tacrolimus antibody-coated paramagnetic microparticles and an acridinium-tacrolimus tracer. The assay was free from hematocrit interference in the range 25%-55% and from interference by extremes of cholesterol, triglycerides, bilirubin, total protein, and uric acid. The total percent of coefficient of variations of the assay were 4.9%-7.6% at 3 ng/mL, 2.9%-4.6% at 8.6 ng/mL, and 3.1%-8.2% at 15.5 ng/mL. Limit of detection was < or =0.5 ng/mL and limit of quantification (LOQ) ranged from 0.69 to 1.07 ng/mL across the 6 sites (based on the upper 95% confidence interval concentrations). The 2007 European Consensus Conference on Tacrolimus Optimization recommended the use of assay methods with an LOQ around 1 ng/mL, based upon the need to measure trough tacrolimus blood concentrations precisely down to 3 ng/mL during low-dose tacrolimus regimens. Tacrolimus International Proficiency Testing Scheme samples were measured by the ARCHITECT immunoassay at 5 sites and showed an average bias of -0.28 to +0.85 ng/mL versus IMx Tacrolimus II immunoassay historical values and -0.21 to +0.68 ng/mL versus liquid chromatography/tandem mass spectrometry (LC-MSMS) Tacrolimus historical values. Method comparison studies were performed with the ARCHITECT Tacrolimus immunoassay on patient specimens with the following results: ARCHITECT Tacrolimus assay versus the Abbott IMx Tacrolimus II immunoassay (4 sites) yielded average biases between -0.94 and +0.26 ng/mL; ARCHITECT assay versus the Dade Dimension Tacrolimus immunoassay (2 sites) yielded average biases of -0.46 and +0.11 ng/mL; and ARCHITECT assay versus LC-MSMS methods at 2 sites yielded average biases of +0.51 and +1.63 ng/mL. Spearman correlation coefficients were >/=0.90 on all method comparisons. The ARCHITECT Tacrolimus assay is a semiautomated, robust, and highly sensitive immunoassay, representing an alternative approach for laboratories not equipped with LC-MSMS, and meets the 1 ng/mL recommendation of LOQ by the European Consensus Conference on Tacrolimus Optimization.
منابع مشابه
Evaluation of the Abbott ARCHITECT HCV Ag Assay
Eight hundred and three samples from a single HCV antibody negative diagnostic population were tested on the Abbott ARCHITECT HCV Ag assay. All samples initially reactive were retested in duplicate. Samples repeatedly reactive were tested for HCV RNA to determine the true status. Samples that were confirmed as positive were excluded from the specificity and negative delta calculations. A total ...
متن کاملThe need for standardization of tacrolimus assays.
BACKGROUND Owing to the lack of an internationally recognized tacrolimus reference material and reference method, current LC-MS and immunoassay test methods used to monitor tacrolimus concentrations in whole blood are not standardized. The aim of this study was to assess the need for tacrolimus assay standardization. METHODS We sent a blinded 40-member whole-blood tacrolimus proficiency panel...
متن کاملMulti-center Performance Evaluations of Tacrolimus and Cyclosporine Electrochemiluminescence Immunoassays in the Asia-Pacific Region
BACKGROUND The immunosuppressant drugs (ISDs), tacrolimus and cyclosporine, are vital for solid organ transplant patients to prevent rejection. However, toxicity is a concern, and absorption is highly variable across patients; therefore, ISD levels need to be precisely monitored. In the Asia-Pacific (APAC) region, tacrolimus and cyclosporine concentrations are typically measured using immunoass...
متن کاملPerformance evaluation and cross-reactivity from insulin analogs with the ARCHITECT insulin assay.
BACKGROUND Insulin measurement is used for the diagnosis of hypoglycemia and for insulin pharmacokinetic evaluations. We assessed the analytical and clinical performance of the ARCHITECT insulin assay, a chemiluminescent immunoassay recently introduced for the ARCHITECT i2000 fully automated immunoassay analyzer (Abbott Laboratories). We also tested whether major insulin analogs cross-reacted w...
متن کاملMultisite analytical evaluation of the Abbott ARCHITECT cyclosporine assay.
The objective of this study was to evaluate the analytical performance of the Abbott ARCHITECT Cyclosporine (CsA) immunoassay in 7 clinical laboratories in comparison to liquid chromatography/tandem mass spectrometry (LC/MS/MS), Abbott TDx, Cobas Integra 800, and the Dade Dimension Xpand immunoassay. The ARCHITECT assay uses a whole blood specimen, a pretreatment step with organic reagents to p...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Therapeutic drug monitoring
دوره 31 2 شماره
صفحات -
تاریخ انتشار 2009