Fracture of the femoral adapter bolt and taper adapter in a modern rotating platform knee arthroplasty
نویسندگان
چکیده
A 58-year-old woman presented with onset of pain associated with a popping sound after a twisting motion 4 years after left total knee revision arthroplasty. She had a complex medical history, including a reported bone cement allergy, and presented to the hospital unable to bear weight. Plain radiographs revealed a broken femoral component, with the femoral metaphyseal sleeve separated from the distal articular component. During surgery, it was observed that the femoral adapter bolt and taper adapter had both fractured. Scanning electron microscopy of the fracture surfaces of the components confirmed that the implant had failed in fatigue, presumably due to high cyclic loads. Failure at this junction has not been described previously. In this type of knee design, we recommend supporting the distal articular component either with bone, augmentation, and/or bone cement to reduce the risk for this mode of failure.
منابع مشابه
Stress Fracture of the Lateral Femoral Condyle afterTotal Knee Arthroplasty
Background: This is to assess one of the rare complications after total knee replacement and to assess risk factors of failure.Methods: 11 patients with varus knee and an average age of 67 years underwent TKA between 2005 and 2013. All patients returned with a sudden sharp knee pain, disability to walk and significant decrease in ROM about 4 to 8 weeks after surgery. Radiographic examination re...
متن کاملPolyethylene Bearing Stress Fractures after Mobile-bearing Cruciate-retaining Total Knee Arthroplasty: Case Reports
Some reported bearing fractures are due to entrapment of the bearing between the femoral component and posterior edge of the tibial base plate in unstable meniscal –bearing design TKA. However, so far no cases of bearing fracture has been reported for the rotating platform or anteroposterior gliding designs of mobile bearing TKA. The authors describe two cases of polyethylene stress fracture af...
متن کاملRemoving a well-fixed femoral sleeve during revision total knee arthroplasty
The following surgical technique describes a case of a 51-year-old man with severe juvenile rheumatoid arthritis that required a 2-stage revision of an infected revision total knee implant. The patient had previously been implanted with a revision rotating platform, constrained condylar device which gained excellent fixation through the use of diaphyseal-engaging stems, and a well-ingrown, full...
متن کاملPosterior cruciate-retaining, rotating-platform total knee arthroplasty: minimum 4-year follow-up study.
Rotating-platform, mobile-bearing total knee arthroplasty was initially developed using a posterior cruciate-sacrificing technique and design. Posterior cruciate-substituting and -retaining, rotating-platform total knee arthroplasty designs were developed later. The purpose of this study was to evaluate a minimum 4-year follow-up of total knee arthroplasty performed with a posterior cruciate-re...
متن کاملRadiologic Assessment of Distal Femur Cutting Angle in Varus Knee Candidates for Total Knee Arthroplasty
Background: In a total knee arthroplasty surgery the goal is to produce 90 degree angle between the knee articular lobe and the mechanical femoral line. Most orthopedic surgeons usually utilize a 5 to 7 degree for distal femoral cutting angle. In this study we will aim at clearing this question, that whether the” five-seven degree” distal femoral cutting angle supposed to be an equable spectrum...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 3 شماره
صفحات -
تاریخ انتشار 2017