Enhancement of Arterial Pressure Pulsatility by Controlling Continuous-Flow Left Ventricular Assist Device Flow Rate in Mock Circulatory System
نویسندگان
چکیده
Continuous-flow left ventricular assist devices (CF-LVADs) generally operate at a constant speed, which reduces pulsatility in the arteries and may lead to complications such as functional changes in the vascular system, gastrointestinal bleeding, or both. The purpose of this study is to increase the arterial pulse pressure and pulsatility by controlling the CF-LVAD flow rate. A MicroMed DeBakey pump was used as the CF-LVAD. A model simulating the flow rate through the aortic valve was used as a reference model to drive the pump. A mock circulation containing two synchronized servomotor-operated piston pumps acting as left and right ventricles was used as a circulatory system. Proportional-integral control was used as the control method. First, the CF-LVAD was operated at a constant speed. With pulsatile-speed CF-LVAD assistance, the pump was driven such that the same mean pump output was generated. Continuous and pulsatile-speed CF-LVAD assistance provided the same mean arterial pressure and flow rate, while the index of pulsatility increased significantly for both arterial pressure and pump flow rate signals under pulsatile speed pump support. This study shows the possibility of improving the pulsatility of CF-LVAD support by regulating pump speed over a cardiac cycle without reducing the overall level of support.
منابع مشابه
Comparison of continuous-flow and pulsatile-flow left ventricular assist devices: is there an advantage to pulsatility?
BACKGROUND Continuous-flow left ventricular assist devices (CFVAD) are currently the most widely used type of mechanical circulatory support as bridge-to-transplant and destination therapy for end-stage congestive heart failure (HF). Compared to the first generation pulsatile-flow left ventricular assist devices (PFVADs), CFVADs have demonstrated improved reliability and durability. However, CF...
متن کاملArterial Pulsatility and Aortic Valve Function under Continuous Flow Left Ventricular Assist
Continuous Flow Left Ventricular Assist Devices (CF-LVADs) generally operate at a constant speed, which causes a decrease in pulse pressure and pulsatility in the arteries and allegedly may lead to late complications such as aortic insufficiency and gastrointestinal bleeding. The purpose of this study is to increase the arterial pulse pressure and pulsatility while obtaining more physiological ...
متن کاملA mathematical model to evaluate control strategies for mechanical circulatory support.
Continuous flow ventricular assist devices (VADs) for mechanical circulatory support (MCS) are generally smaller and believed to be more reliable than pulsatile VADs. However, regarding continuous flow, there are concerns about the decreased pulsatility and ventricular unloading. Moreover, pulsatile VADs offer a wider range in control strategies. For this reason, we used a computer model to eva...
متن کاملBeneficial aspects of real time flow measurements for the management of acute right ventricular heart failure following continuous flow ventricular assist device implantation
BACKGROUND Optimal management of acute right heart failure following the implantation of a left ventricular assist device requires a reliable estimation of left ventricular preload and contractility. This is possible by real-time pump blood flow measurements. CLINICAL CASE We performed implantation of a continuous flow left ventricular assist device in a 66 years old female patient with an en...
متن کاملA Mock Circulatory System to Assess the Performance of Continuous-Flow Left Ventricular Assist Devices (LVADs): Does Axial Flow Unload Better Than Centrifugal LVAD?
Hemodynamic performances comparisons between different types of left ventricular assist devices (LVADs) remain difficult in a clinical context. The aim of this study was to create an experimental model to assess and compare two types of LVAD under hemodynamic conditions that simulated physical effort and pulmonary hypertension. An experimental mock circulatory system was created to simulate the...