Mode III cleavage of a coin-shaped titanium implant in bone: effect of friction and crack propagation.
نویسندگان
چکیده
Endosseous cementless implants are widely used in orthopaedic, maxillofacial and oral surgery. However, failures are still observed and remain difficult to anticipate as remodelling phenomena at the bone-implant interface are poorly understood. The assessment of the biomechanical strength of the bone-implant interface may improve the understanding of the osseointegration process. An experimental approach based on a mode III cleavage mechanical device aims at understanding the behaviour of a planar bone-implant interface submitted to torsional loading. To do so, coin-shaped titanium implants were inserted on the tibiae of a New Zealand white rabbit for seven weeks. After the sacrifice, mode III cleavage experiments were performed on bone samples. An analytical model was developed to understand the debonding process of the bone-implant interface. The model allowed to assess the values of different parameters related to bone tissue at the vicinity of the implant with the additional assumption that bone adhesion occurs over around 70% of the implant surface, which is confirmed by microscopy images. The approach allows to estimate different quantities related to the bone-implant interface such as: torsional stiffness (around 20.5 N m rad(-1)), shear modulus (around 240 MPa), maximal torsional loading (around 0.056 N.m), mode III fracture energy (around 77.5 N m(-1)) and stress intensity factor (0.27 MPa m(1/2)). This study paves the way for the use of mode III cleavage testing for the investigation of torsional loading strength of the bone-implant interface, which might help for the development and optimization of implant biomaterial, surface treatment and medical treatment investigations.
منابع مشابه
Fatigue and corrosion fatigue properties of Ti-6Al-4V implant grade titanium alloy in Ringer solution
Nowadays modification of metallic biomaterials which are used as implants for bone and hard tissues replacement is considered as an important subject. In the current study, corrosion fatigue properties of Ti-6Al-4V alloy investigated via Rotating-Bending standard test method and then, the results compared with the fatigue properties of the specimens tested in the same conditions. Scanning elect...
متن کاملThe use of a coin shaped implant for direct in situ measurement of attachment strength for osseointegrating biomaterial surfaces.
Most animal models currently used to study the retention of implants in bone are influenced by shear forces introduced during the retention test. This is mainly due to the implant design, which most often are cylindrical, conical or threaded. In these models interlocking between bone and implant surface will increase the effect of genuine bone bonding and thus give a false positive outcome. The...
متن کاملThe axisymmetric computational study of a femoral component to analysis the effect of titanium alloy and diameter variation.
This work presents a numerical approach in order to predict the influence of implant material stiffness in a femoral component design when submitted in compression. The implant success depends on the transferred load to the neighboring bone. The finite element method can be used to analysis the stress and strain distribution in the femoral component allowing to improve the implant selection. Fo...
متن کاملEffect of micro-roughness produced by TiO2 blasting--tensile testing of bone attachment by using coin-shaped implants.
The aim of the present study was to examine bone response to micro-rough titanium implants. Forty coin-shaped implants were divided into eight groups according to their surface roughness. The first group had electropolished surfaces. The surfaces of implant groups 2-8 were blasted with TiO2 particles with incremental grain sizes ranging from 7.5-12.5 to 270-330 microns. Five implants from each ...
متن کاملExtended finite element simulation of crack propagation in cracked Brazilian disc
The cracked Brazilian disc (CBD) specimen is widely used in order to determine mode-I/II and mixed-mode fracture toughness of a rock medium. In this study, the stress intensity factor (SIF) on the crack-tip in this specimen is calculated for various geometrical crack conditions using the extended-finite element method (X-FEM). This method is based upon the finite element method (FEM). In this m...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of the mechanical behavior of biomedical materials
دوره 8 شماره
صفحات -
تاریخ انتشار 2012