Quantifying the effect of air gap, depth, and range shifter thickness on TPS dosimetric accuracy in superficial PBS proton therapy
نویسندگان
چکیده
This study quantifies the dosimetric accuracy of a commercial treatment planning system as functions of treatment depth, air gap, and range shifter thickness for superficial pencil beam scanning proton therapy treatments. The RayStation 6 pencil beam and Monte Carlo dose engines were each used to calculate the dose distributions for a single treatment plan with varying range shifter air gaps. Central axis dose values extracted from each of the calculated plans were compared to dose values measured with a calibrated PTW Markus chamber at various depths in RW3 solid water. Dose was measured at 12 depths, ranging from the surface to 5 cm, for each of the 18 different air gaps, which ranged from 0.5 to 28 cm. TPS dosimetric accuracy, defined as the ratio of calculated dose relative to the measured dose, was plotted as functions of depth and air gap for the pencil beam and Monte Carlo dose algorithms. The accuracy of the TPS pencil beam dose algorithm was found to be clinically unacceptable at depths shallower than 3 cm with air gaps wider than 10 cm, and increased range shifter thickness only added to the dosimetric inaccuracy of the pencil beam algorithm. Each configuration calculated with Monte Carlo was determined to be clinically acceptable. Further comparisons of the Monte Carlo dose algorithm to the measured spread-out Bragg Peaks of multiple fields used during machine commissioning verified the dosimetric accuracy of Monte Carlo in a variety of beam energies and field sizes. Discrepancies between measured and TPS calculated dose values can mainly be attributed to the ability (or lack thereof) of the TPS pencil beam dose algorithm to properly model secondary proton scatter generated in the range shifter.
منابع مشابه
Using field size factors to characterize the in-air fluence of a proton machine with a range shifter
INTRODUCTION The range shifter (RS) is used to treat shallow tumors for a proton pencil beam scanning system (PBS). Adding RS certainly complicates the commissioning of the treatment planning system (TPS) because the spot sizes are significantly enlarged with RS. In this work, we present an efficient method to configure a commercial TPS for a PBS system with a fixed RS. METHODS By combining a...
متن کاملWater Equivalent Thickness Analysis of Immobilization Devices for Clinical Implementation in Proton Therapy
Immobilization devices can impact not only the inter- and intra-fraction motion of the patient, but also the range uncertainty of the treatment beam in proton therapy. In order to limit additional range uncertainty, the water equivalent thickness (WET) of the immobilization device needs to be well known and accurately reflected in the calculations by the treatment planning system (TPS). The met...
متن کاملAccuracy Evaluation of Isogray TPS Dose Calculations in Symmetric and Asymmetric Fields of the Elekta Compact Linear Accelerator
Introduction: Radiation therapy is one of the most important methods of cancer treatment. Radiation therapy uses symmetric and asymmetric fields in which the radiation dose distribution is different. To calculate the dose distribution, computer treatment planning systems are used which must have acceptable calculations accuracy. The aim of this study was to investigate the accuracy of the dose ...
متن کاملEvaluation of the range shifter model for proton pencil‐beam scanning for the Eclipse v.11 treatment planning system
Existing proton therapy pencil-beam scanning (PBS) systems have limitations on the minimum range to which a patient can be treated. This limitation arises from practical considerations, such as beam current intensity, layer spacing, and delivery time. The range shifter (RS) - a slab of stopping material inserted between the nozzle and the patient - is used to reduce the residual range of the in...
متن کاملEffects of Defining Realistic Compositions of the Ocular Melanoma on Proton Therapy
Background: Recent studies in eye plaque brachytherapy have shown a considerable difference between the dosimetric results using water phantom and a model of human eye containing realistic materials. In spite of this fact, there is a lack of simulation studies based on such a model in proton therapy literatures. In the presented work, the effect of utilizing an eye model with ocular media on pr...
متن کامل