Acetabular revisions using porous tantalum components: A retrospective study with 5-10 years follow-up
نویسندگان
چکیده
AIM To evaluate the clinical and X-ray results of acetabular components and tantalum augments in prosthetic hip revisions. METHODS Fifty-eight hip prostheses with primary failure of the acetabular component were reviewed with tantalum implants. The clinical records and X-rays of these cases were retrospectively reviewed. Bone defect evaluations were based on preoperative CT scans and classified according to Paprosky criteria of Radiolucent lines and periprosthetic gaps; implant mobilization and osteolysis were evaluated by X-ray. An ad hoc database was created and statistical analyses were performed with SPSS software (IBM SPSS Statistics for Windows, version 23.0). Statistical analyses were carried out using the Student's t test for independent and paired samples. A P value of < 0.05 was considered statistically significant and cumulative survival was calculated by the Kaplan-Meier method. RESULTS The mean follow-up was 87.6 ± 25.6 mo (range 3-120 mo). 25 cases (43.1%) were classified as minor defects, and 33 cases (56.9%) as major defects. The preoperative HHS rating improved significantly from a mean of 40.7 ± 6.1 (range: 29-53) before revision, to a mean of 85.8 ± 6.1 (range: 70-94) at the end of the follow-up (Student's t test for paired samples: P < 0.001). Considering HHS only at the end of follow-up, no statistically significant difference was observed between patients with a major or minor defect (Student's t test for independent samples: P > 0.05). Radiolucent lines were found in 4 implants (6.9%). Postoperative acetabular gaps were observed in 5 hips (8.6%). No signs of implant mobilization or areas of periprosthetic osteolysis were found in the x-rays at the final follow-up. Only 3 implants failed: 1 case of infection and 2 cases of instability. Defined as the end-point, cumulative survival at 10 years was 95% (for all reasons) and 100% for aseptic loosening of the acetabular component. CONCLUSION The medium-term use of prosthetic tantalum components in prosthetic hip revisions is safe and effective in a wide variety of acetabular bone defects.
منابع مشابه
Multiple Acetabular Revisions in THA - Poor Outcome Despite Maximum Effort
A consecutive series of 52 acetabular revisions was evaluated retrospectively. Inclusion criteria for all patients were at least one former exchange of the acetabular component. Reconstruction was performed with reliable techniques and implants other than extensively porous coated device (e.g. tantalum). The mean follow up was 5.63 (0.01-14.05) years. Cumulative survival at 14.05 years with rem...
متن کاملThe Trabecular Metal Monoblock acetabular component in patients with high congenital hip dislocation: a prospective study.
Between November 1997 and December 2000 we performed 27 total hip replacements in 22 patients with high congenital dislocation of the hip using porous tantalum monoblock acetabular components implanted in the true acetabular bed. Clinical and radiological evaluation was performed at regular intervals for a mean of 10.2 years (8.5 to 12). The mean Harris Hip Score improved from 48.3 (15 to 65) p...
متن کاملRevision Total Hip Arthroplasty Using Tantalum Augment in Patients with Paprosky III or IV Acetabular Bone Defects: A Minimum 2-year Follow Up Study
PURPOSE The purpose of this study is to report the short-term outcomes of revision total hip arthroplasty (THA) using tantalum augments in patients with severe acetabular bone defects. MATERIALS AND METHODS We retrospectively analyzed 15 revision THAs performed in 15 patients using tantalum augments between June 2010 and December 2013. Acetabular bone defects were Paprosky type IIIA in 7 hips...
متن کاملA retrospective cohort study of 207 acetabular revisions
© 2016 The Author(s). Published by Taylor & Francis on behalf of the Nordic Orthopedic Federation. This is an Open Access article distributed under the terms of the Creative Commons Attribution-Non-Commercial License (https://creativecommons.org/licenses/by-nc/3.0) DOI 10.1080/17453674.2016.1248315 Background and purpose — Porous tantalum cups have been introduced as an alternative to various r...
متن کاملEarly results of trabecular metal augment for acetabular reconstruction in revision hip arthroplasty.
The management of acetabular bone defects presents a challenge in revision total hip arthroplasty (THA). The aim of this study was to assess the early clinical and radiological outcome of revision of acetabular components using trabecular metal cups and augments for acetabular reconstruction. The study included 18 consecutive patients with failed acetabular components after total hip arthroplas...
متن کامل