Ventilator-disconnect and death: a case study and a safety device.

نویسنده

  • William S Lowery
چکیده

Deaths and injuries related to accidental ventilator disconnection occur in complex ways. A death related to accidental ventilator disconnection is presented from a subacute ventilator facility, and corrective action is discussed. Single-limb ventilators are now equipped with a flow-bypass coupling that prevents patient-side occlusion during disconnect and therefore prevents low-pressure alarm malfunction. Physicians and respiratory therapists can consider this device to prevent partial or total occlusion of the ventilator tracheostomy adapter, thus allowing a low-pressure alarm in the event of a disconnect.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Identification of Human Errors in the Use of Neonatal Ventilator Device by Predictive Use Error Analysis method

Introduction: Nowadays, accidents are regarded as a main risk factor for both human and economic of countries. The triggerring cause of most accident is human error. In the healthcare setting, human errors can lead to injuries and even death of patients and damage the reputation of healthcare staff. Human errors in healthcares can occur during various activities including diagnosis of a disease...

متن کامل

Ventilator with a capnograph smart sensors

Today, the world has witnessed increasing development in various fields, including industry and medicine. One of the best ways to achieve these objectives is the integration of different sciences including electronics, robotics and medicine. Ventilator set that provides respiratory function for patients get smarter and more capable than older generations, during mechanical ventilation to preven...

متن کامل

Frequency of device-associated infections in intensive care units

Background: Health care associated infections (HAIs) are a class of infections that infect patients during hospital admissions and receive medical services. These infections occurs within 48 to 72 hours of admission and up to 6 weeks after discharge. Surveillance of device-associated infections (DAIs) in intensive care units (ICUs) is substantial in planning healthcare strategies. This study wa...

متن کامل

Improving Patient Safety With X-Ray and Anesthesia Machine Ventilator Synchronization: A Medical Device Interoperability Case Study

When a x-ray image is needed during surgery, clinicians may stop the anesthesia machine ventilator while the exposure is made. If the ventilator is not restarted promptly, the patient may experience severe complications. This paper explores the interconnection of a ventilator and simulated x-ray into a prototype plug-and-play medical device system. This work assists ongoing interoperability fra...

متن کامل

Ecmo in a diabetic patient with cerebellar infarction and ARDS- A case report

Background: Extra corporeal membrane oxygenation (ECMO) has been considered as a crucial modality for the management of acute respiratory distress syndrome (ARDS). Since ECMO is considered to be an invasive modality and may be associated with some complications, there are strict indications for its use. Here we report successful treatment of a case of severe ARDS with Veno-Venous ECMO. Case pre...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Respiratory care

دوره 55 6  شماره 

صفحات  -

تاریخ انتشار 2010