Results with commercial radioassay kits compared with microbiological assay of folate in serum and whole-blood.
نویسندگان
چکیده
We compared results with three commercial folate radioassay kits [Bio-Rad, New England Nuclear (NEN), and RIA Products] with those by microbiological assay for more than 200 samples of human serum and whole blood. All but one kit (NEN) compared favorably with the microbiological assay for serum samples, although there were notable diagnostic discrepancies. Two kits (NEN and Bio-Rad) were tested on whole-blood samples; both yielded values significantly higher than those by microbiological assay. The frequency distributions of erythrocyte folate data differed strikingly between the two kits; the NEN method yielded a much narrower range of normal values than did either the Bio-Rad or the microbiological assay. Radioassay kits appear to be suitable diagnostic agents for serum folate, if the behavior of a particular kit is investigated thoroughly before its routine use. However, the diagnostic value of radioassays of erythrocyte folate needs to be validated.
منابع مشابه
Analytical evaluation of a folate radioassay: application to serum and erythrocyte measurements.
Many commercial folate radioassay kits have proven to be unsatisfactory in the past. It has been found that the Schwarz/Mann Iodine-125 folate radioassay kit permits a rapid measurement of folic acid requiring only a single incubation at pH 9.3. At this pH, the relative affinity of pteroylglutamic acid (FA) and N-5-methyltetrahydrofolic acid (MTFA) for the milk binder is the same. Additionally,...
متن کاملComparison of radioassay and microbiological assay for serum folate, with clinical assessment of discrepant results.
Folate assays by use of radiolabeled folate provide obvious practical advantages over the standard microbiological assay, but remain incompletely tested. We therefore compared results for 415 sera with a kit involving 3H-labeled folate and the Lactobacillus casei microbiological method. We examined the patients' data when there were discrepancies between the two methods. Although the correlatio...
متن کاملMeasurement of red cell folate with 75Se-selenofolate radioassay.
75Se-selenofolate radioassay was adapted for the purpose of red cell folate measurement. The results of assay of 124 whole blood samples by Lactobacillus casei microbiological assay and by radioassay using the adapted selenofolate method were concordant (r = 0.8). Red cell folate levels by the radioassay were below 170 microgram/1 in people who had haematological features of folate depletion. T...
متن کاملComparison of commercial kits for radioimmunoassay: III. Radioassay of serum folate.
We evaluated six commercial radioassay kits in which tritium or iodine tracers are used to measure serum folate. We assessed the adequacy of the literature supplied with the kits, within-run precision, reagent stability, analytical recovery of added pteroylglutamic acid and N-methyltetrahydrofolic acid, results of analysis of patients' samples, and minimum detection limits. There were serious d...
متن کاملRadioassay of serum folate.
We describe a radioassay procedure for serum folate, in which commercially available crystalline 3lactoglobulirt is used as the folate-binding agent. The folate binders present in serum are eliminated by heating serum in a boiling water bath for 15 mm to denature the proteins. Radioassayable serum folate is shown to be stable under these conditions, and the results for folate compare well with ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 24 12 شماره
صفحات -
تاریخ انتشار 1978