Numerical investigations of MRI RF field induced heating for external fixation devices
نویسندگان
چکیده
BACKGROUND The magnetic resonance imaging (MRI) radio frequency (RF) field induced heating on external fixation devices can be very high in the vicinity of device screws. Such induced RF heating is related to device constructs, device placements, as well as the device insertion depth into human subjects. In this study, computational modeling is performed to determine factors associated with such induced heating. METHODS Numerical modeling, based on the finite-difference time-domain (FDTD) method, is used to evaluate the temperature rises near external device screw tips inside the ASTM phantom for both 1.5-T and 3-T MRI systems. The modeling approach consists of 1) the development of RF coils for 1.5-T and 3-T, 2) the electromagnetic simulations of energy deposition near the screw tips of external fixation devices, and 3) the thermal simulations of temperature rises near the tips of these devices. RESULTS It is found that changing insertion depth and screw spacing could largely affect the heating of these devices. In 1.5-T MRI system, smaller insertion depth and larger pin spacing will lead to higher temperature rise. However, for 3-T MRI system, the relation is not very clear when insertion depth is larger than 5 cm or when pin spacing became larger than 20 cm. The effect of connection bar material on device heating is also studied and the heating mechanism of the device is analysed. CONCLUSIONS Numerical simulation is used to study RF heating for external fixation devices in both 1.5-T and 3-T MRI coils. Typically, shallower insertion depth and larger pin spacing with conductive bar lead to higher RF heating. The heating mechanism is explained using induced current along the device and power decay inside ASTM phantom.
منابع مشابه
MR safety assessment of potential RF heating from cranial fixation plates at 7 T.
PURPOSE The increasing number of clinically oriented MRI studies at 7 T motivates the safety assessment of implants, since many 7 T research sites conservatively exclude all subjects with metallic implants, regardless of type or location. The purpose of this study was to investigate potential RF-induced heating during a 7 T MRI scan using a self-built transmit/receive RF coil in patients with i...
متن کاملMeasuring RF-induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leads.
Radiofrequency (RF)-related heating of cardiac pacemaker leads is a serious concern in magnetic resonance imaging (MRI). Recent investigations suggest such heating to be strongly dependent on an implant's position within the surrounding medium, but this issue is currently poorly understood. In this study, phantom measurements of the RF-induced electric currents inside a pacemaker lead were perf...
متن کاملIn vivo and post-mortem measurements of radio frequency induced heating during MRI of pigs implanted with vascular stents
Background Radio Frequency (RF) induced heating associated with MRI is an increasing safety concern as more patients are implanted with cardiovascular devices. A standard test method (ASTM F2182) is used across the device industry to measure RF heating, but this method does not account for cooling mechanisms such as blood flow and perfusion. We hypothesize that in vivo conditions will significa...
متن کاملEvaluation of Radio Frequency-Induced Heating near Fixed Orthodontic Appliances during MRI Examination at 3 Tesla
The purpose of this paper is to assess the added risk pose by radio frequency (RF) induced heating near fixed orthodontic appliances in patients undergoing MRI examination at 3 Tesla. Three of the most commonly used types of fixed orthodontic appliances: Palatal expander (Hyrax) Lingual Bar and Multibracket-/-band braces were investigated using a head phantom and positioned to match the in-vivo...
متن کاملComparison of temperature measurements of pacemaker leads in a 1.0T high field open MRI and a 1.5T classic cylindrical MRI: initial results
Introduction As of today, the use of MRI procedures on patients with implanted cardiac pacemakers is prohibited due to safety issues. The implants can interact with the RF fields of the MRI device. The most hazardous effect is heating at the tip of the lead, less dangerous are sensing errors and malfunctions of the devices, because they disappear completely after the procedure. The majority of ...
متن کامل