Necrobiosis and T-lymphocyte infiltration in retrieved aseptically loosened metal-on-polyethylene arthroplasties
نویسندگان
چکیده
BACKGROUND AND PURPOSE Soft tissue necrobiosis and T-lymphocyte infiltration within the periprosthetic soft tissue have been linked to a suggested hypersensitivity reaction of the delayed-type following the metal-on-metal arthroplasty. While we observed both synovial necrobiosis and lymphocyte infiltrates in synovial tissues with failed metal-on-polyethylene prostheses, we hypothesized that both findings are unspecific for metal-on-metal bearing coupes. Thus, we wished to quantify the extent of necrobiosis and the amount of T-lymphocyte infiltration in aseptically loosened metal-on-polyethylene prostheses. MATERIALS AND METHODS We analyzed 28 consecutive synovial biopsy specimens obtained at revision surgery of aseptically loosened metal-on-polyethylene prostheses (19 hips and 9 knees) and quantified both the extent of necrobiosis vertically and the density of CD3+, CD4+, and CD8+ lymphocytes within the joint capsular tissue. We excluded patients with inflammatory skeletal disease or with a history of metal hypersensitivity. RESULTS We found necrobiosis in 23 of 28 cases and it was most often connected with the superficial portions of the synovium. Necrobiosis of deeper tissues was seen in 8 specimens and it was strongly associated with superficial necrobiosis. While CD3+ lymphocytes were detected in each biopsy, 4 cases with more than 300 CD3+ lymphocytes were identified in the group of 26 cases that presented with more than 100 CD3+ lymphocytes within one high-power field. 16 cases with more than 100 CD3+ lymphocytes also showed concomitant superficial necrobiosis of the synovium. In the inflammatory infiltration of periprosthetic synovium, CD8+ lymphocytes predominated over CD4+ cells. INTERPRETATION Synovial necrobiosis and infiltration of T-lymphocytes are common findings in tissues around aseptically loosened metal-on-polyethylene arthroplasty in patients without a clinically suspected metal hypersensitivity reaction. Thus, neither necrobiosis nor infiltration of T-lymphocytes should be considered to be specific for failed metal-on-metal bearings or metal hypersensitivity reaction.
منابع مشابه
Proliferation of the synovial lining cell layer in suggested metal hypersensitivity.
Synovial tissues in joints with prostheses display characteristic morphological changes in cases with aseptic failure, particularly macrophage infiltration. Since proliferation of the synovial lining cell layer represents a feature characteristic of autoimmune joint diseases, the possibility of morphological changes of the synovial lining cell layer in periprosthetic tissues was investigated. S...
متن کاملRheumatoid nodules at the cement-bone interface in revision arthroplasty of rheumatoid patients.
We studied the histology of the cement-bone interface and the regenerated synovium in ten aseptically loosened arthroplasties in rheumatoid patients. In two patients we found rheumatoid nodules at the cement-bone interface and marked lymphoplasmacytic infiltration in another three. Failed joints in osteoarthritic patients did not show these changes, and it seems likely that the presence of abun...
متن کاملCD3+ lymphocytosis in the peri-implant membrane of 222 loosened joint endoprostheses depends on the tribological pairing
Background and purpose - The most frequent cause of arthroplasty failure is aseptic loosening-often induced by particles. Abrasion material triggers inflammatory reactions with lymphocytic infiltration and the formation of synovial-like interface membranes (SLIM) in the bone-implant interface. We analyzed CD3 quantities in SLIM depending on articulating materials and possible influences of prov...
متن کاملMetal-on-metal total hip arthroplasty: known and unknown side effects.
The indications for hip replacement, especially in young patients with end-stage degenerative arthritis of the hip, have increased during the past 20 years. Because of the accelerated risk of polyethylene wear and subsequent prosthetic loosening due to osteolysis, new interest has focused on the metal-on-metal articulation couple. Failures of first-generation metal-on-metal total hip arthroplas...
متن کاملMetal release and metal allergy after total hip replacement with resurfacing versus conventional hybrid prosthesis
BACKGROUND Metal-on-metal (MOM) total hip arthroplasties were reintroduced because of the problems with osteolysis and aseptic loosening related to polyethylene wear of early metal-on-polyethylene (MOP) arthroplasties. The volumetric wear rate has been greatly reduced with MOM arthroplasties; however, because of nano-size wear particles, the absolute number has been greatly increased. Thus, a s...
متن کامل