Validation of the Omron HEM 742 blood pressure monitoring device in adolescents.
نویسندگان
چکیده
BACKGROUND Accurate blood pressure measurement is fundamental for scientific investigation or clinical decision-making. In this sense, it is important to verify the values provided by electronic devices. OBJECTIVE To validate the Omron HEM 742 blood pressure monitoring device in adolescents according to criteria suggested by the British Hypertension Society. METHODS A total of 150 adolescents aged between 10 and 16 years participated in the study. The automated Omron HEM 742 monitor was connected in Y to the mercury column auscultatory device, then three simultaneous measurements were taken, and the differences between the readings of the two devices were calculated. The intraclass correlation coefficient and Bland-Altman plot (agreement) were used to verify the relationship between both devices. Specificity and sensitivity of the device were determined by using the ROC curve. RESULTS The comparison between the measurements showed an equal to or lower than 5mmHg difference in 67.3% of the systolic values, and 69.3% of the diastolic values; an equal to or lower than 10mmHg difference occurred in 87.3% and 90.6% of the systolic and diastolic values, respectively; an equal to or lower than 15mmHg difference was found in 96.6% of the systolic values and 97.3% of the diastolic values. These findings are consistent with a grade A according to the British Hypertension Society protocol. A marked consonance was observed between the values obtained by the automated monitor and this device was proven to be capable of identifying the presence or absence of high blood pressure. CONCLUSION The Omron HEM 742 monitor was proved valid for blood pressure measurement in adolescents according to the criteria suggested by the British Hypertension Society.
منابع مشابه
Validation of the Omron HEM-907 device for blood pressure measurement.
BACKGROUND The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol. DESIGN The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which ...
متن کاملValidation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT).
BACKGROUND Two electronic devices for self-measurement of blood pressure - a brachial monitor, the Omron M6, and a wrist monitor, the Omron R7 - were evaluated in two separate studies according to the International Protocol of the European Society of Hypertension. DESIGN The International Validation Protocol is divided into two phases: the first phase is performed on 15 selected participants ...
متن کاملValidation of the Omron HEM-7320-LA, upper arm blood pressure monitor with Intelli Wrap Technology Cuff HEM-FL1 for self-measurement and clinic use according to the European Society of Hypertension International Protocol revision 2010 in the Mexican population
OBJECTIVE The aim of this study was to determine the accuracy of the Omron HEM-7320-LA with Intelli Wrap technology cuff HEM-FL1 for self-measurement and clinic blood pressure (BP) measurement according to the European Society of Hypertension International Protocol revision 2010. PARTICIPANTS AND METHODS The evaluation was performed in 39 individuals. The mean age of the participants was 47.9...
متن کاملValidation study of an automated wrist monitor, omron model HEM-608, compared with the standard methods for blood pressure measurement.
OBJECTIVE - The aim of our study was to assess the profile of a wrist monitor, the Omron Model HEM-608, compared with the indirect method for blood pressure measurement. METHODS - Our study population consisted of 100 subjects, 29 being normotensive and 71 being hypertensive. Participants had their blood pressure checked 8 times with alternate techniques, 4 by the indirect method and 4 with the...
متن کاملValidation of Omron RS8, RS6, and RS3 home blood pressure monitoring devices, in accordance with the European Society of Hypertension International Protocol revision 2010
BACKGROUND Allowing patients to measure their blood pressure at home is recognized as being of clinical value. However, it is not known how often these measurements are taken correctly. Blood pressure monitors for home use fall into two types based on the position of the cuff, ie, at the upper arm or the wrist. The latter is particularly convenient, as measurements can be taken fully clothed. T...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Arquivos brasileiros de cardiologia
دوره 92 1 شماره
صفحات -
تاریخ انتشار 2009