Airway scope and Macintosh laryngoscope for tracheal intubation in patients lying on the ground.

نویسندگان

  • Ryu Komatsu
  • Kotoe Kamata
  • Daniel I Sessler
  • Makoto Ozaki
چکیده

BACKGROUND Direct laryngoscopy of a patient lying on the ground is difficult because the intubator's head is far above the head of the patient, making alignment of the intubator's visual axis with the patient's tracheal axis difficult. The Airway Scope is a laryngoscope designed to facilitate tracheal intubation without requiring alignment of the oral, pharyngeal, and tracheal axes. We thus tested the hypothesis that intubation with the Airway Scope is faster than with the Macintosh laryngoscope in subjects lying on the ground. METHODS Adult surgical patients were enrolled. After anesthesia induction, direct laryngoscopy was performed and airway characteristics noted. Patients were randomly assigned to tracheal intubation by either the Airway Scope (n = 50) or the Macintosh laryngoscope (n = 50). The intubator performed tracheal intubation from a table positioned at the same height as that of the operating table, thus simulating intubating on the ground. An unblinded observer recorded overall intubation success rate, time required for intubation, the number of attempts required for successful intubation, and airway complications related to intubation. Of these, the primary end point was time required for intubation. RESULTS Overall intubation success rates were 98% with the Airway Scope and 100% with the Macintosh laryngoscope. Intubation was 17 s faster with the Airway Scope (mean, 18 (SD, 4) seconds) versus the Macintosh laryngoscope (35 (16) seconds). The number of intubation attempts was similar with each device. The incidences of airway complications were similar, with no hypoxia (Spo(2) <95%) occurring in either group. CONCLUSIONS Both the Airway Scope and the Macintosh laryngoscope offer high success rates in adequately prepared paralyzed patients lying supine at ground level in the hands of a skilled practitioner. However, the Airway Scope facilitated faster tracheal intubation.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the success rate and quality of teaching of tracheal intubation to medical students by using video laryngoscope

Background: Training of airway management and tracheal intubation skill in emergencies for resuscitation of patients are basic skills for medical students. However, the success rate of beginners in this skill is low. Video laryngoscopes are new devices that can increase the success of endotracheal intubation training. Methods: This clinical trial was conducted with 30 medical students who came...

متن کامل

Comparison of the Incidence of Sore Throat and Hemodynamic Changes in ‎Airway Management by Macintosh, Glideoscope Laryngoscope and ‎Laryngeal Mask Airway Techniques: Randomized Controlled Trial

Background: Studies have shown that postoperative sore throat is the most common complication of tracheal intubation, However, the actual incidence and extent of it are not well documented. Objective: The aim of this study was to evaluate and compare the incidence of sore throat and hemodynamic Changes in Macintosh, Glidescope Laryngoscope and laryngeal mask airway  techniques in elective surg...

متن کامل

Tracheal intubation in daylight and in the dark: a randomised comparison of the Airway Scope, Airtraq, and Macintosh laryngoscope in a manikin.

SUMMARY Fifteen anaesthetists attempted to intubate the trachea of a manikin lying supine on the ground using the Airway Scope, Airtraq or Macintosh laryngoscope in three simulated conditions: (1) in room light; (2) in the dark and (3) in daylight. The main outcome measure was the time to ventilate the lungs after successful intubation; the secondary outcome was the success rate of ventilation ...

متن کامل

The usefulness of the McGrath MAC laryngoscope in comparison with Airwayscope and Macintosh laryngoscope during routine nasotracheal intubation: a randomaized controlled trial

BACKGROUND McGrath MAC video laryngoscope offers excellent laryngosopic views and increases the success rate of orotracheal intubation in some cases. The aim of this study was to determine the usefulness of McGrath MAC for routine nasotracheal intubation by comparing McGrath MAC with Airway scope and Macintosh laryngoscope. METHODS A total of 60 adult patients with ASA physical status class 1...

متن کامل

Airway scope and gum elastic bougie with Macintosh laryngoscope for tracheal intubation in patients with simulated restricted neck mobility.

BACKGROUND /st> Airway Scope is a new videolaryngoscope which requires less cervical movement during intubation than direct laryngoscopy. Thus, in patients wearing a rigid cervical collar, we compared the efficacy of the Airway Scope and the gum elastic bougie with Macintosh laryngoscope during tracheal intubation. METHODS /st> Anaesthesia was induced with propofol, fentanyl, and rocuronium. ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Anesthesia and analgesia

دوره 111 2  شماره 

صفحات  -

تاریخ انتشار 2010