Comparative performances analysis of neonatal ventilators
نویسندگان
چکیده
BACKGROUND Mechanical ventilation is a therapeutic action for newborns with respiratory diseases but may have side effects. Correct equipment knowledge and training may limit human errors. We aimed to test different neonatal mechanical ventilators' performances by an acquisition module (a commercial pressure sensor plus an isolated chamber and a dedicated software). METHODS The differences (ΔP) between peak pressure values and end-expiration pressure were investigated for each ventilator. We focused on discrepancies among measured and imposed pressure data. A statistical analysis was performed. RESULTS We investigated the measured/imposed ΔP relation. The ΔP do not reveal univocal trends related to ventilation setting parameters and the data distributions were non-Gaussian. CONCLUSIONS Measured ΔP represent a significant parameter in newborns' ventilation, due to the typical small volumes. The investigated ventilators showed different tendencies. Therefore, a deep specific knowledge of the intensive care devices is mandatory for caregivers to correctly exploit their operating principles.
منابع مشابه
New generation neonatal high frequency ventilators: effect of oscillatory frequency and working principles on performance.
BACKGROUND Several new generation neonatal ventilators that incorporate conventional as well as high frequency ventilation (HFOV) have appeared on the market. Most of them offer the possibility to use HFOV in a volume-targeted mode, despite absence of any preclinical data. With a bench test, we evaluated the performances of 4 new neonatal HFOV devices and compared them to the SensorMedics HFOV ...
متن کاملNeonatal and adult ICU ventilators to provide ventilation in neonates, infants, and children: a bench model study.
BACKGROUND Using a bench test model, we investigated the hypothesis that neonatal and/or adult ventilators equipped with neonatal/pediatric modes currently do not reliably administer pressure support (PS) in neonatal or pediatric patient groups in either the absence or presence of air leaks. METHODS PS was evaluated in 4 neonatal and 6 adult ventilators using a bench model to evaluate trigger...
متن کاملPerformance characteristics of five new anesthesia ventilators and four intensive care ventilators in pressure-support mode: a comparative bench study.
BACKGROUND During the past few years, many manufacturers have introduced new modes of ventilation in anesthesia ventilators, especially partial-pressure modalities. The current bench test study was designed to compare triggering and pressurization of five new anesthesia ventilators with four intensive care unit ventilators. METHODS Ventilators were connected to a two-compartment lung model. O...
متن کاملBench-test comparison of 26 emergency and transport ventilators
INTRODUCTION Numerous emergency and transport ventilators are commercialized and new generations arise constantly. The aim of this study was to evaluate a large panel of ventilators to allow clinicians to choose a device, taking into account their specificities of use. METHODS This experimental bench-test took into account general characteristics and technical performances. Performances were ...
متن کاملA Nationwide Census of ICU Capacity and Admissions in Mongolia
In Mongolia, a Central Asian lower-middle income country, intensive care medicine is an under-resourced and-developed medical specialty. The burden of critical illness and capacity of intensive care unit (ICU) services in the country is unknown. In this nationwide census, we collected data on adult and pediatric/neonatal ICU capacities and the number of ICU admissions in 2014. All hospitals reg...