Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing.
نویسندگان
چکیده
OBJECTIVES To determine the effects of low, normal, and high hematocrit levels on glucose meter measurements and to assess the clinical risks of hematocrit errors. DESIGN Changes in glucose measurements between low and high hematocrit levels were calculated to determine hematocrit effects. The differences between glucose measured with meters and with a plasma glucose method (YSI 2300) also were compared. SETTING Six hand-held glucose meters were assessed in vitro at low (19.1%), normal (38.5%), and high (58.3%) hematocrit levels, and at 6 glucose concentrations ranging from 2.06 mmol/L (37.1 mg/dL) to 30.24 mmol/L (544.7 mg/dL). RESULTS Most systems, regardless of the reference to which they were calibrated, demonstrated positive bias at lower hematocrit levels and negative bias at higher hematocrit levels. Low, normal, and high hematocrit levels progressively lowered Precision G and Precision QID glucose measurements. Hematocrit effects on the other systems were more dependent on the glucose concentration. Overall, Accu-Chek Comfort Curve showed the least sensitivity to hematocrit changes, except at the lowest glucose concentration. CONCLUSIONS We strongly recommend that clinical professionals choose glucose systems carefully and interpret glucose measurements with extreme caution when the patient's hematocrit value changes, particularly if there is a simultaneous change in glucose level.
منابع مشابه
Screening for hypoglycemia at the bedside in the neonatal intensive care unit (NICU) with the Abbott PCx glucose meter
BACKGROUND Point of care (POC) glucose meters are routinely used as a screening tool for hypoglycemia in a neonatal setting. Glucose meters however, lack the same accuracy as laboratory instruments for glucose measurement. In this study we investigated potential reasons for this inaccuracy and established a cut off value for confirmatory testing. METHODS In this prospective study, all patient...
متن کاملPerformance of a new meter designed for assisted monitoring of blood glucose and point-of-care testing.
BACKGROUND Blood glucose (BG) meters used for assisted monitoring of blood glucose (AMBG) require different attributes compared with meters designed for home use. These include safety considerations (i.e., minimized risk of blood-borne pathogen transmission), capability for testing multiple blood sample types, and enhanced performance specifications. The OneTouch® Verio™Pro+ BG meter is designe...
متن کاملDynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices.
BACKGROUND It has been demonstrated that dynamic electrochemistry can be used to correct blood glucose measurement results for potentially interfering conditions, such as humidity, hematocrit (HCT) variations, and ascorbic acid. The purpose of this laboratory investigation was to assess the potential influence of hematocrit variations on a variety of blood glucose meters applying different meas...
متن کاملComparison of point-of-care and laboratory glucose analysis in critically ill patients.
BACKGROUND Blood for point-of-care analysis of glucose levels is often obtained from different sources (fingerstick, arterial or central venous catheter). OBJECTIVES To examine agreement between point-of-care and laboratory glucose values and to determine effects of hematocrit, serum carbon dioxide, and mean arterial pressure on the accuracy of point-of-care values. METHODS Point-of-care va...
متن کاملSuboptimal Performance of Blood Glucose Meters in an Antenatal Diabetes Clinic
OBJECTIVE The objective of this study was to evaluate the performance of blood glucose meters in diabetes associated with pregnancy (DP). RESEARCH DESIGN AND METHODS Finger-prick blood glucose levels measured using six different glucose meters on 102 patients with DP attending an antenatal clinic were compared with laboratory plasma glucose results. HbA(1c) and hematocrit were also measured. ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Archives of pathology & laboratory medicine
دوره 124 8 شماره
صفحات -
تاریخ انتشار 2000