A comparative analysis of OTF, NPS, and DQE in energy integrating and photon counting digital x-ray detectors.

نویسندگان

  • Raymond J Acciavatti
  • Andrew D A Maidment
چکیده

PURPOSE One of the benefits of photon counting (PC) detectors over energy integrating (EI) detectors is the absence of many additive noise sources, such as electronic noise and secondary quantum noise. The purpose of this work is to demonstrate that thresholding voltage gains to detect individual x rays actually generates an unexpected source of white noise in photon counters. METHODS To distinguish the two detector types, their point spread function (PSF) is interpreted differently. The PSF of the energy integrating detector is treated as a weighting function for counting x rays, while the PSF of the photon counting detector is interpreted as a probability. Although this model ignores some subtleties of real imaging systems, such as scatter and the energy-dependent amplification of secondary quanta in indirect-converting detectors, it is useful for demonstrating fundamental differences between the two detector types. From first principles, the optical transfer function (OTF) is calculated as the continuous Fourier transform of the PSF, the noise power spectra (NPS) is determined by the discrete space Fourier transform (DSFT) of the autocovariance of signal intensity, and the detective quantum efficiency (DQE) is found from combined knowledge of the OTF and NPS. To illustrate the calculation of the transfer functions, the PSF is modeled as the convolution of a Gaussian with the product of rect functions. The Gaussian reflects the blurring of the x-ray converter, while the rect functions model the sampling of the detector. RESULTS The transfer functions are first calculated assuming outside noise sources such as electronic noise and secondary quantum noise are negligible. It is demonstrated that while OTF is the same for two detector types possessing an equivalent PSF, a frequency-independent (i.e., "white") difference in their NPS exists such that NPS(PC) > or = NPS(EI) and hence DQE(PC) < or = DQE(EI). The necessary and sufficient condition for equality is that the PSF is a binary function given as zero or unity everywhere. In analyzing the model detector with Gaussian blurring, the difference in NPS and DQE between the two detector types is found to increase with the blurring of the x-ray converter. Ultimately, the expression for the additive white noise of the photon counter is compared against the expression for electronic noise and secondary quantum noise in an energy integrator. Thus, a method is provided to determine the average secondary quanta that the energy integrator must produce for each x ray to have superior DQE to a photon counter with the same PSF. CONCLUSIONS This article develops analytical models of OTF, NPS, and DQE for energy integrating and photon counting digital x-ray detectors. While many subtleties of real imaging systems have not been modeled, this work is illustrative in demonstrating an additive source of white noise in photon counting detectors which has not yet been described in the literature. One benefit of this analysis is a framework for determining the average secondary quanta that an energy integrating detector must produce for each x ray to have superior DQE to competing photon counting technology.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Detective quantum efficiency dependence on x-ray energy weighting in mammography.

An evaluation of the dependence of detective quantum efficiency (DQE) on the incident energy spectrum has been made for mammography. The DQE dependence on the energy spectrum has been evaluated for energy-integrating detectors, photon-counting detectors, and detectors that measure the energy of each photon. To isolate the effect of the x-ray energy spectrum the detector has been assumed to be i...

متن کامل

Modeling the frequency-dependent detective quantum efficiency of photon-counting x-ray detectors.

PURPOSE To find a simple model for the frequency-dependent detective quantum efficiency (DQE) of photon-counting detectors in the low flux limit. METHODS Formula for the spatial cross-talk, the noise power spectrum and the DQE of a photon-counting detector working at a given threshold are derived. Parameters are probabilities for types of events like single counts in the central pixel, double...

متن کامل

An experimental comparison of detector performance for direct and indirect digital radiography systems.

Current flat-panel detectors either directly convert x-ray energy to electronic charge or use indirect conversion with an intermediate optical process. The purpose of this work was to compare direct and indirect detectors in terms of their modulation transfer function (MTF), noise power spectrum (NPS), and detective quantum efficiency (DQE). Measurements were made on three flat-panel detectors,...

متن کامل

X-ray Imaging Detector Using Silicon Field Emission Tip Array Energy Conversion

Existing x-ray detecting technologies can be classified as either photon-counting (digital) or integrating (analog). Multi-wire proportional counters (MVVPC) are digital detectors that have advantages such as high quantum sensitivity, high dynamic range, and high intrinsic energy resolution, and have a wide range of applications. However, most photon-counting detectors rely on gas conversion of...

متن کامل

Calculation of OTF, NPS, and DQE for Oblique X-Ray Incidence on Turbid Granular Phosphors

Digital breast tomosynthesis (DBT) is an imaging modality in which tomographic sections of the breast are generated from a limited range of x-ray tube angles. One drawback of DBT is resolution loss in the oblique projection images. The purpose of this work is to extend Swank’s formulation of the transfer functions of turbid granular phosphors to oblique x-ray incidence, using the diffusion appr...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Medical physics

دوره 37 12  شماره 

صفحات  -

تاریخ انتشار 2010