Collision indicator charts for gantry‐couch position combinations for Varian linacs
نویسنده
چکیده
The use of non-coplanar radiation fields can potentially lead to collisions between the gantry and the couch or patient. The collisions are often not realized until the plan is finished and the fields are checked on the machine, or even later when the patient is already on the table. This paper presents an easy method of gauging if a collision is likely between the gantry and couch or patient during treatment planning. The method involves creating a chart of allowable gantry and couch combinations. The charts contain curves on a polar graph of the gantry and couch angle "plane". The curves display the limits of collisions for each gantry and couch combination for vertical couch positions 10, 15 and 20 cm below isocenter and for couch lateral positions of -10, 0, +10 cm, covering the majority of couch positions encountered in patient treatments. All combinations in the region within the curves (containing the origin) are valid, while all combinations outside the curves will result in a collision. The data for the charts are collected from measurements of the gantry angle that just clears each couch angle. The patient presence was modeled by placing a stereotactic body frame on the top of the couch. Separate charts were created for couch angles between 0° and 90° and between 360° and 270° over all gantry angles. The graphs are easy to create, implement, and use in the clinic and help reduce the time, complications, and uncertainties of planning with non-coplanar fields.
منابع مشابه
Collision indicator charts for gantry‐couch position combinations for Siemens ONCOR and Elekta Infinity linacs
Noncoplanar radiation fields from a linear accelerator can be used to deliver radiation dose distributions that are superior to those delivered using coplanar radiation fields. Noncoplanar radiation field arrangements are especially valuable when delivering stereotactic body radiation therapy (SBRT). Noncoplanar radiation fields, however, are geometrically more challenging to deliver than copla...
متن کاملAbstractID: 9678 Title: An analytical approach to prevent gantry-couch collision for linac-based radiosurgery
ID: 9678 Title: An analytical approach to prevent gantry-couch collision for linac-based radiosurgery Gantry-couch collision is a serious concern for the treatment planning of the linac-based stereotactic radiosurgery (SRS). The ability to detect collision at the planning time removes the need for backup plans and preserves the useful beam angles that would be deemed unsafe and discarded otherw...
متن کاملAn EPID-based method for comprehensive verification of gantry, EPID and the MLC carriage positional accuracy in Varian linacs during arc treatments
BACKGROUND In modern radiotherapy, it is crucial to monitor the performance of all linac components including gantry, collimation system and electronic portal imaging device (EPID) during arc deliveries. In this study, a simple EPID-based measurement method has been introduced in conjunction with an algorithm to investigate the stability of these systems during arc treatments with the aim of en...
متن کاملEvaluation of IsoCal geometric calibration system for Varian linacs equipped with on‐board imager and electronic portal imaging device imaging systems
The purpose of this study is to evaluate the accuracy and reproducibility of the IsoCal geometric calibration system for kilovoltage (kV) and megavoltage (MV) imagers on Varian C-series linear accelerators (linacs). IsoCal calibration starts by imaging a phantom and collimator plate using MV images with different collimator angles, as well as MV and kV images at different gantry angles. The sof...
متن کاملTH-AB-BRB-01: Trajectory Modulated Arc Therapy: Application to Partial Breast Irradiation.
Current state-of-the art digital C-arm medical linear accelerators are capable of delivering radiation treatments with high level of automation, which affords coordinated motions of gantry, couch, and multileaf collimator (MLC) with dose rate modulations. The new machine capacity has shown the potential to bring substantially improved radiation dosimetry and/or delivery efficiency to many chall...