Avoiding hyperoxemia during neonatal resuscitation: time to response of different SpO2 monitors

نویسندگان

  • Hernando Baquero
  • Ramiro Alviz
  • Armando Castillo
  • Fredy Neira
  • Augusto Sola
چکیده

AIM To assess the time to obtain reliable oxygen saturation readings by different pulse oximeters during neonatal resuscitation in the delivery room or NICU. METHODS Prospective study comparing three different pulse oximeters: Masimo Radical-7 compared simultaneously with Ohmeda Biox 3700 or with Nellcor N395, in newborn infants who required resuscitation. Members of the research team placed the sensors for each of the pulse oximeters being compared simultaneously, one sensor on each foot of the same baby. Care provided routinely, without interference by the research team. The time elapsed until a reliable SpO2 was obtained was recorded using a digital chronometer. Statistical comparisons included chi-square and student's T-test. RESULTS Thirty-two infants were enrolled; median gestational age 32 weeks. Seventeen paired measurements were made with the Radical-7 and Biox 3700; mean time to a stable reading was 20.2±7 sec for the Radical-7 and 74.2±12 sec for the Biox 3700 (p=0.02). The Radical-7 and the N- 395 were paired on 15 infants; the times to obtain a stable reading were 20.9±4 sec and 67.3±12 sec, respectively (p=0.03). CONCLUSION The time to a reliable reading obtained simultaneously in neonatal critical situations differs by the type of the pulse oximeter used, being significantly faster with Masimo Signal Extraction Technology. This may permit for better adjustments of inspired oxygen, aiding in the prevention of damage caused by unnecessary exposure to high or low oxygen.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Oxygen in neonatal anesthesia: friend or foe?

PURPOSE OF REVIEW Clinical practices in oxygen administration are in need of change based on the increasing understanding of oxygen toxicity. Hypoxemia is due to many pathophysiological causes; avoiding hypoxemia is an important objective during neonatal anesthesia. Nevertheless, the only known cause for hyperoxemia is the excess and unnecessary administration of oxygen by healthcare providers....

متن کامل

Hyperoxemia – too much of a good thing?

While avoiding hypoxemia has long been a goal of critical care practitioners, less attention has been paid to the potential for excessive oxygenation. Interest has mounted recently in understanding the clinical effects of hyperoxemia during critical illness, in particular its impact following cardiac arrest. In this issue of Critical Care, Dell'Anna and colleagues review available animal and hu...

متن کامل

Automated versus Manual Oxygen Control with Different Saturation Targets and Modes of Respiratory Support in Preterm Infants.

OBJECTIVE To determine the efficacy and safety of automated adjustment of the fraction of inspired oxygen (FiO2) in maintaining arterial oxygen saturation (SpO2) within a higher (91%-95%) and a lower (89%-93%) target range in preterm infants. STUDY DESIGN Eighty preterm infants (gestational age [median]: 26 weeks, age [median] 18 days) on noninvasive (n = 50) and invasive (n = 30) respiratory...

متن کامل

Feasibility and reliability of an automated controller of inspired oxygen concentration during mechanical ventilation

INTRODUCTION Hypoxemia and high fractions of inspired oxygen (FiO2) are concerns in critically ill patients. An automated FiO2 controller based on continuous oxygen saturation (SpO2) measurement was tested. Two different SpO2-FiO2 feedback open loops, designed to react differently based on the level of hypoxemia, were compared. The results of the FiO2 controller were also compared with a histor...

متن کامل

Assessment of Delivery Room Resuscitation with Different Levels and Its Related Factors in Preterm Neonates

Background: There are many known risk factors related to maternal or neonatal problems which can predict the need for resuscitation. In this study, we evaluated the resuscitation process of preterm neonates and analyzed the impact of different risk factors on the level of resuscitation required in the patients.Methods: This cross-sectional descriptive study was conducted on inborn preterm...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 100  شماره 

صفحات  -

تاریخ انتشار 2011