Clinical evaluation of Eastman Kodak's Ektachem 400 Analyzer.
نویسندگان
چکیده
We evaluated the performance of the Kodak Ektachem 400 Analyzer in a 16-week clinical trial. We assessed four potentiometric tests and nine colorimetric tests for precision and correlation with results obtained some other commonly used instruments (Technicon SMA II and C800 System, Du Pont aca II, and Baker CentrifiChem). The comparison was favorable for all tests except albumin, sodium, and carbon dioxide. The reagents are stable. Little refrigerator storage space is needed. The computer system and the modular component design of the Ektachem 400 Analyzer make operation, troubleshooting, and preventive maintenance relatively easy.
منابع مشابه
Extended clinical trial and evaluation of glucose determination with the Eastman Kodak Ektachem GLU/BUN Analyzer.
We followed the "abbreviated precision protocol" of the National Committee for Clinical Laboratory Standards for the evaluation of precision, accuracy, and carryover in analyses for glucose with the "Ektachem." We analyzed 760 clinical samples by this technique, by the FDA Proposed Class Standard glucose reference method, and with the Beckman System I GLU/BUN Analyzer. Precision and accuracy we...
متن کاملFactitiously high sodium activities on the Ektachem 400 owing to interferences by high gamma-globulin concentrations.
Sera from patients with above-normal gamma-globulin concentrations give sodium values as much as 10% higher when measured in the Kodak Ektachem 400 analyzer than results obtained with the Beckman Astra-8 analyzer. Results obtained with the Astra were in agreement with flame photometry. A similar relative bias was observed for potassium measurement in the Ektachem. This discrepancy was corrected...
متن کاملGeneration 04 sodium slides and generation 04 reference fluid evaluated for use with the Ektachem 400 Analyzer.
Evaluation of a new generation of sodium slides and reference fluid for use with the Kodak Ektachem 400 Analyzer demonstrated improved performance in generation 04 over the previous generation 03 slide for sodium determinations. A decrease in between-replicate imprecision and a correction of the "serum" assigned values of the calibrators used to calibrate the slide used with the reference fluid...
متن کاملInterferences with potentiometry of CO2 in the Ektachem 400 Analyzer.
We have investigated the Kodak Ektachem 400 Analyzer procedure for CO2 for interferences from benzyl alcohol, benzoic acid, and several compounds structurally similar to benzoic acid. Benzoic acid in plasma, at concentrations found in neonates intoxicated with benzyl alcohol, caused a large increase in the results for CO2, as did substantially above-normal concentrations of certain fatty acids ...
متن کاملMeasurement of urinary chloride with the Kodak Ektachem 400.
Urinary chloride can be measured simply, precisely, and accurately with the Ektachem 400 clinical chemistry analyzer. Because the coated film technique used in this instrument requires the presence of a protein matrix, we added a protein-based calibrator to the urine samples to overcome the low protein concentration in these specimens. Its advantages over a coulometric titration method (Astra a...