A New Generation Prothrombin Time Method for INR
نویسندگان
چکیده
Prothrombin time (PT) is the leading test for monitoring oral anticoagulation therapy (OAT). We sought to determine INR taking into account only active coagulation factors FII, FVII and FX without inhibition in patient plasmas and calibrator kits.We measured PT using a combined thromboplastin reagent. The calculation was based on a new PT method, which measures active coagulation factors (F II, F VII, FX) and corrects the errors caused by inactive coagulation factors.On this basis, an INR result with and without inhibition for individual patient samples was also calculated and applied to 200 plasma samples obtained from OAT patients. Conspicuous variation in inhibition between the four calibration kits was noted. The kinetics of this inhibition was closest to a noncompetitive pattern.The need of correction for INRs of single patients increases with higher INRs. At the same level of patient INRs the coagulation inhibiton varies markedly.It has been known that different thromboplastin reagents possess variable sensitivities, but this may depend on sensitivity in inactive coagulation factors. PT methods today measure the sum of active coagulation factors and inhibition of inactive coagulation factors. ISI calibrators contain variable amounts of inactive coagulation factors, which renders harmonisation of INR results.Application of the Acf-PT (INR(Acf)) presented in this work develops the PT methodology to measure the true coagulation activity in vivo for patient warfarin therapy without inhibition. INR(Inh) can evidently also be used for the diagnostics and follow-up of certain liver diseases.
منابع مشابه
A sensitivity comparison of the Quick and Owren prothrombin time methods in oral anticoagulant therapy
Prothrombin time (PT) is the leading test for monitoring oral anticoagulation therapy (OAT). According to the World Health Organization recommendation, International Normalized Ratio (INR) results obtained from the same patient samples with the major PT methods (Quick and Owren) should be the same when the therapeutic range is the same. In our study blood samples were obtained from 207 OAT pati...
متن کاملPoint-of-care monitoring of oral anticoagulation therapy in children. Comparison of the CoaguChek XS system with venous INR and venous INR using an International Reference Thromboplastin preparation (rTF/95).
Point-of-care (POC) monitoring of oral anticoagulation has been widely adopted in both paediatric and adult patients. A new POC system, the CoaguChek XS has recently been developed to measure the international normalised ratio (INR) and may offer significant advantages. The CoaguChek XS utilises a new method of electrochemical clot detection based on thrombin generation. This system has not bee...
متن کاملInterpretation of serial measurements of international normalized ratio for prothrombin times in monitoring oral anticoagulant therapy.
Despite careful monitoring of oral anticoagulant treatment (OAT), some international normalized ratio (INR) for prothrombin time values will fall outside the therapeutic range. Considerable changes in serial INR results from OAT patients may be caused by random fluctuation alone, and, for statistical reasons, a fraction of the INR values will fall outside therapeutic range and interfere with do...
متن کاملInternational normalized ratio for prothrombin times in patients taking oral anticoagulants: critical difference and probability of significant change in consecutive measurements.
To determine when a change in serial measurements of prothrombin time in patients receiving oral anticoagulant therapy (ACT) is a statistically significant biological change necessitating dose adjustment, one must know the size of the "critical difference" in statistical terms (i.e., probabilities). In a cohort of 32 ACT patients at pharmacological steady-state, we studied the within-subject to...
متن کاملA New-Generation Method for Quick and Owren PT
Oral anticoagulant therapy (OAT) calls for continuous control by prothrombin time (PT) test, as the therapeutic range in INR units is very narrow. Warfarin (or coumarin) inhibits coagulation factor synthesis in the liver, but at the same time inactive coagulation factors are formed. The aim here was to measure "active coagulation factors" and inhibition in calibrator kits and patient plasmas by...
متن کامل