High accuracy Kerr effect measurement technique

نویسنده

  • D. P. Shelton
چکیده

Apparatus and techniques are demonstrated which allow gas phase dc electro-optic Kerr effect measurements to be made with 0.1% absolute accuracy and nanoradian sensitivity. Measurements of the electric-field-induced birefringence of a gas are calibrated against a liquid Kerr cell, which is in turn calibrated absolutely in situ. Some of the problems addressed in this work are control of stray birefringence in order to obtain an extinction ratio below lo-“, a simple method for calibration of space charge effects in the CS, reference Kerr cell, construction of electrodes with accurately calculable fields, and an investigation of the sources of systematic errors and drifts and means to eliminate them.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Propagation Technique for Ultrashort Pulses II: Numerical Methods to Solve the Pulse Propagation Equation

We presented the numerical technique to approximately solve the pulse propagation equation. Two efficient methods for this problem, the Split-Step Fourier and the fourth order Runge-Kutta methods are considered. Their high accuracy are shown by comparison with analytical solutions in some particular situations. Our numerical experiments are implemented for soliton propagation and interacting hi...

متن کامل

Employing dual frequency phase sensitive demodulation technique to improve the accuracy of voltage measurement in magnetic induction tomography and designing a labratoary prototype

Magnetic induction tomography (MIT) is a promising modality for noninvasive imaging due to its contactless technology. Being a non-contact safe imaging technique, MIT has been an appropriate method in compare to other electrical tomography. In this imaging method, a primary magnetic field is applied by excitation coils to induce eddy currents in the material to be studied and a secondary magnet...

متن کامل

Measurement of pure Kerr nonlinearity in GaN thin films at 800 nm by means of eclipsing Z-scan experiments

We report the measurement of Kerr nonlinearity of thin films of GaN by using the eclipsing Z-scan technique. The measurement was performed using 100 fs pulses at 800 nm. We measured a pure refractive signal, whose associated nonlinear Kerr coefficient was n2 = (−7.3 ± 0.4)× 10−14 cm2 W−1.

متن کامل

Keer electro-optic measurements in liquid dielectrics

Kerr electro-optic technique has been used to measure the electric field distribution in high voltage stressed dielectric liquids, where the difference between refractive indices for light polarized parallel and perpendicular to the local electric field is a function of the electric field intensity. For transformer oil, the most widely-used insulating liquids in power apparatus and high voltage...

متن کامل

Comparing the Accuracy of Cone Beam Computed Tomography,Digital Intraoral Radiography and Conventional Intraoral Radiography in the Measurement of Periodontal Bone Defects

Background and Aim: Cone beam computed tomography (CBCT) produces high-quality data in periodontal diagnosis and treatment planning. The aim of this study was to compare the accuracy of CBCT with intraoral digital and conventional radiography in the measurement of periodontal bone defects.  Methods and Materials: In this diagnostic research, two hundred and eighteen artificial osseous d...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 1999