A novel acetabular alignment guide for THR using selective anatomic landmarks on the pelvis.
نویسندگان
چکیده
This paper presents a novel approach for acetabular alignment during the implant of a prosthetic hip joint in a natural pelvis. The alignment instrument uses selective anatomic bony landmarks on the pelvis, which are accessible in surgery, to guide the placement of the acetabular component in the appropriate orientation. A closed form solution, involving both a forward and reverse analysis, is presented to relate the parameters of the device with the abduction and anteversion angles. Using mathematical models, this device should allow the surgeon to place the acetabular component with an orientation between 10.9 degrees and 19.1 degrees anteversion and 35.7 degrees and 44.3 degrees abduction with 95% confidence in a male/left specimen for the commonly accepted target of 15 degrees anteversion and 40 degrees abduction. This device is currently being used successfully by one of the authors in THR surgery.
منابع مشابه
Effects of Spinopelvic Balance on Acetabular Version in Total Hip Arthroplasty
The orientation of acetabular component is influenced by pelvic tilt, body position and individual variation in pelvic parameters. Most adverse events attribute to improper positioning of the acetabular component in the functional position. There is evidence that the orientation of the pelvis changes from the supine to standing position, consequently changing the orientation of acetabular cup v...
متن کاملUsing acetabular fossa as a guide for anticipated inclination of uncemented cup in total hip replacement.
Positions of acetabular implant generally are considered to be major causative factors of dislocation. Accurate and consistent achievement of the preoperatively anticipated orientation of the acetabular cup is a great challenge in total hip replacement (THR). In the present study, we investigated the surgical application of acetabular fossa as a guide for anticipated inclination of uncemented c...
متن کاملLocalization of Pelvic Anatomical Coordinate System Using US/Atlas Registration for Total Hip Replacement
In Total Hip Replacement (THR) procedures, misalignment of the acetabular component can lead to dislocation and impingement. For the successful alignment of acetabular component, precise estimation of pelvic anatomical coordinate system is necessary. Conventional navigation systems use CT scan or fluoroscopy, or involve implanted bone fiducials or invasive probing of bony landmarks to locate th...
متن کاملUltrasound Based Localization of Pelvic Anatomical Coordinate System
INTRODUCTION: In Total Hip Replacement, correct alignment of the acetabular component requires precise localization of the pelvic anatomical coordinate system [1, 2]. Typically, computed tomography and fluoroscopy have been used, in conjunction with implanted fiducials and invasive probing of bony landmarks. In this work, we propose an ultrasound based approach that exploits prior knowledge abo...
متن کاملDistances between bony landmarks and adjacent nerves: anatomical factors that may influence retractor placement in total hip replacement surgery.
BACKGROUND Retractor placement is a leading cause of intraoperative nerve injury during total hip replacement (THR) surgery. The sciatic nerve, femoral nerve, and superior gluteal nerve are most commonly affected. This study aimed to identify the distances from bony landmarks in the hip to the adjacent nerves on magnetic resonance imaging (MRI) and the associations between anatomical factors an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of biomechanics
دوره 38 9 شماره
صفحات -
تاریخ انتشار 2005