Six-year experience with the Delta III reverse shoulder prosthesis.

نویسندگان

  • Federico Alberto Grassi
  • Luigi Murena
  • Federico Valli
  • Riccardo Alberio
چکیده

PURPOSE To report the clinical and radiographic results and complications of the Delta III reverse prosthesis. METHODS 24 women and 2 men aged 62 to 84 (mean, 75) years underwent total shoulder replacement using the Delta III reverse prosthesis. Patient diagnoses were massive rotator cuff tear (n=20), disabling sequelae of proximal humeral fractures (n=3), and failure of an unconstrained arthroplasty (n=3). Clinical and functional results were assessed using the Constant scale. Active range of motion (ROM) was measured. Scapular notching and radiolucent lines around the humeral component were evaluated using radiographs. Patient satisfaction of the treatment was evaluated by a direct interview. RESULTS 23 patients were followed up for 26 to 84 (mean, 42) months. Two patients had loosening of the glenoid component (at 6 months and 5 years) and underwent revision surgery. There were no instances of infection, instability, or acromial fracture. Only active elevation improved significantly after surgery, as did both the absolute and adjusted Constant scores. 12 patients were completely pain-free, 9 complained of slight pain, and one of moderate pain. The severity of scapular notching progressed with time. 15 patients were satisfied with the treatment, 6 were partially satisfied, and 2 were not satisfied. CONCLUSION The Delta III prosthesis restores shoulder function but has biomechanical limits. Its use should be limited to elderly patients with severe impairment of the glenohumeral joint. Scapular notching is a main concern for the long-term survival of the implant.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Midterm Results of the Delta Xtend Reverse Shoulder System: A Five-Year Outcome Study

Background: The purpose of this study was to examine the mid-term functional outcomes, radiographic results,and revision rates of patients treated with the Delta Xtend Reverse Shoulder System for both primary and revisionarthroplasty indications.Methods: A retrospective review was conducted of records for all individuals who underwent a reverse shoulderarthroplasty using the Delta Xtend Reverse...

متن کامل

The reverse shoulder prosthesis (Delta III) in acute shoulder fractures: technical considerations with respect to stability.

The reverse shoulder prosthesis reverses the relationship between the scapular and humeral component, resulting in a mechanical advantage as the deltoid muscle is able to compensate for the rotator cuff deficiency. Based on this mechanical advantage, the reverse shoulder prosthesis has become an accepted alternative for the treatment of complex proximal humeral fractures. The purpose of this ar...

متن کامل

Reverse shoulder arthroplasty: does reduced medialisation improve radiological and clinical results?

The purpose of this retrospective study was to assess the effect of the reduced medialisation of the Arrow reverse shoulder prosthesis on short-term clinical and radiological results. The clinical and radiological results of 47 Delta III reverse prostheses and 49 Arrow reverse prostheses were retrospectively compared at a minimum of twelve months follow-up. There was a significant increase (p <...

متن کامل

Revision of shoulder replacement with a reversed shoulder prosthesis (Delta III): report of five cases.

The authors report on their experience in revision total shoulder arthroplasty using the Delta III shoulder prosthesis, a reversed constrained prosthesis which is known to give good functional results in the rotator cuff-deficient shoulder. It was implanted in patients with a failed total or hemishoulder prosthesis. The Constant-Murley score was used to assess the functional outcome. The prelim...

متن کامل

Comparison of bone removed with reverse total shoulder arthroplasty.

INTRODUCTION A computer model quantified glenoid base - plate surface contact area and the amount of humeral and glenoid bone removed by three different reverse shoulder prosthesis designs when implanted with various techniques. METHODS The computer model quantified differences in glenoid baseplate contact area and the cortical and cancellous humeral and glenoid bone removed to implant the 36...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of orthopaedic surgery

دوره 17 2  شماره 

صفحات  -

تاریخ انتشار 2009