Multipoint pacing via a quadripolar left-ventricular lead: preliminary results from the Italian registry on multipoint left-ventricular pacing in cardiac resynchronization therapy (IRON-MPP)

نویسندگان

  • Giovanni B. Forleo
  • Luca Santini
  • Massimo Giammaria
  • Domenico Potenza
  • Antonio Curnis
  • Vito Calabrese
  • Danilo Ricciardi
  • Carlo D'agostino
  • Pasquale Notarstefano
  • Valentina Ribatti
  • Giovanni Morani
  • Massimo Mantica
  • Luigi Di Biase
  • Emanuele Bertaglia
  • Leonardo Calò
  • Francesco Zanon
چکیده

Aims This registry was created to describe the experience of 76 Italian centres with a large cohort of recipients of multipoint pacing (MPP) capable cardiac resynchronization therapy (CRT) devices. Methods and results A total of 507 patients in whom these devices had been successfully implanted were enrolled between August 2013 and May 2015. We analysed: (i) current clinical practices for the management of such patients, and (ii) the impact of MPP on heart failure clinical composite response and on the absolute change in ejection fraction (EF) at 6 months. Multipoint pacing was programmed to 'ON' in 46% of patients before discharge. Methods of optimizing MPP programming were most commonly based on either the greatest narrowing of the QRS complex (38%) or the electrical delays between the electrodes (34%). Clinical and echocardiographic follow-up data were evaluated in 232 patients. These patients were divided into two groups according to whether MPP was programmed to 'ON' (n = 94) or 'OFF' (n = 138) at the time of discharge. At 6 months, EF was significantly higher in the MPP group than in the biventricular-pacing group (39.1 ± 9.6 vs. 34.7 ± 7.6%; P < 0.001). Even after adjustments, early MPP activation remained an independent predictor of absolute increase in LVEF of ≥5% (odds ratio 2.5; P = 0.001). At 6 months, an improvement in clinical composite score was recorded in a greater proportion of patients with MPP-ON than in controls (56 vs. 38%; P = 0.009). On comparing optimal MPP and conventional vectors, QRS was also seen to have decreased significantly (P < 0.001). Conclusion This study provides information that is essential in order to deal with the expected increase in the number of patients receiving MPP devices in the coming years. The results revealed different practices among centres, and establishing the optimal programming that can maximize the benefit of MPP remains a challenging issue. Compared with conventional CRT, MPP improved clinical status and resulted in an additional increase in EF. Clinical Trial Registration http://www.clinicaltrial.gov/. Unique identifier: NCT02606071.

منابع مشابه

Quadripolar Leads in Cardiac€Resynchronization Therapy

Ma Despite the benefit of cardiac resynchronization therapy (CRT) in patients with heart failure and conduction delay, a considerable number of patients do not respond substantially. Left ventricular lead position is an important factor in response, restricted by the patient’s specific anatomy and local pathophysiology. Quadripolar leads could enhance response to CRT, offering 4 pacing location...

متن کامل

Multipoint Pacing versus conventional ICD in Patients with a Narrow QRS complex (MPP Narrow QRS trial): study protocol for a pilot randomized controlled trial

BACKGROUND Despite an intensive search for predictors of the response to cardiac resynchronization therapy (CRT), the QRS duration remains the simplest and most robust predictor of a positive response. QRS duration of ≥ 130 ms is considered to be a prerequisite for CRT; however, some studies have shown that CRT may also be effective in heart failure (HF) patients with a narrow QRS (<130 ms). Si...

متن کامل

Intentional anodal capture of a left ventricular quadripolar lead enhances resynchronization equally with multipoint pacing

Figure 1 Changes in the 12-lead electrocardiograms (ECGs) during biventricular pacing with different left ventricular (LV) pacing configurations. A: Extended bipolar pacing between LV1 and right ventricular coil. B: Extended bipolar pacing between LV4 and right ventricular coil. C: LV bipolar pacing between LV1 (anode) and LV4 (cathode) with an output of 4.0 V/0.4 ms. The similar morphology to ...

متن کامل

Cardiac Resynchronization Therapy Delivered Via a Multipolar Left Ventricular Lead is Associated with Reduced Mortality and Elimination of Phrenic Nerve Stimulation: Long‐Term Follow‐Up from a Multicenter Registry

INTRODUCTION Cardiac resynchronization therapy (CRT) using quadripolar left ventricular (LV) leads provides more pacing vectors compared to bipolar leads. This may avoid phrenic nerve stimulation (PNS) and allow optimal lead placement to maximize biventricular pacing. However, a long-term improvement in patient outcome has yet to be demonstrated. METHODS A total of 721 consecutive patients wi...

متن کامل

Quadripolar Left Ventricular Lead in a Patient with CRT-D Does Not Overcome Phrenic Nerve Stimulation

Effective cardiac resynchronization therapy (CRT) requires an accurate atrio-biventricular pacing system. The innovative Quartet lead is a quadripolar, over-the-wire left ventricular lead with four electrodes and has recently been designed to provide more options and greater control in pacing vector selection. A lead with multiple pacing electrodes is a potential alternative to physical adjustm...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

متن کامل
عنوان ژورنال:

دوره 19  شماره 

صفحات  -