Fixation and Wear With Contemporary Acetabular Components and Cross-Linked Polyethylene at 10-Years in Patients Aged 50 and Under.

نویسندگان

  • Justin J Greiner
  • John J Callaghan
  • Nicholas A Bedard
  • Steve S Liu
  • Yubo Gao
  • Devon D Goetz
چکیده

We evaluated the mid-term performance of a moderately cross-linked polyethylene THA bearing in younger patients. One hundred consecutive THAs performed in patients 50 years of age and under using the same cementless acetabular component and moderately cross-linked polyethylene were evaluated at minimum 10 year follow-up. At final follow-up 75 patients (89 hips) were living, 7 patients (7 hips) were deceased and 4 patients (4 hips) were lost to follow-up. Average Tegner and UCLA scores were 3.7 and 5.8, respectively. Average steps per year were 1.84 million. The average linear wear rate was 0.05 mm/y and steady state wear rate was 0.033 mm/y. There was no detectable osteolysis. Minimum 10 year follow-up demonstrated durable fixation and reduced wear (P<0.001) when compared to our historical controls.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Results of Primary Total Hip Arthroplasty Using 36 mm Femoral Heads on 1st Generation Highly Cross Linked Polyethylene in Patients 50 Years and Less with Minimum Five Year Follow-up

PURPOSE We evaluated the clinical and radiographic midterm results of primary total hip arthroplasty (THA) using a 36 mm diameter femoral head on 1st generation highly cross-linked polyethylene (HXLPE) in patients 50 years and less with minimum five year follow-up. MATERIALS AND METHODS We retrospectively reviewed 31 patients (41 hips) aged 50 years and less underwent primary THA with a 36 mm...

متن کامل

Hip Resurfacing Using Highly Cross-linked Polyethylene: Prospective Study Results at 8.5 Years.

BACKGROUND Hip resurfacing is an option to consider when treating younger, more active patients. Advantages over total hip arthroplasty include a more normal gait and a lower incidence of thigh pain. METHODS In this prospective study, 190 hip resurfacing procedures (164 participants) were performed using a cobalt-chromium femoral component and a cementless acetabular cup with a 3.8-mm highly ...

متن کامل

Cross-linked compared with historical polyethylene in THA: an 8-year clinical study.

UNLABELLED Wear particle-induced osteolysis is a major cause of aseptic loosening in THA. Increasing wear resistance of polyethylene (PE) occurs by increasing the cross-link density and early reports document low wear rates with such implants. To confirm longer-term reductions in wear we compared cross-linked polyethylene (irradiation in nitrogen, annealing) with historical polyethylene (irradi...

متن کامل

Fixation, polyethylene wear, and pelvic osteolysis in primary total hip replacement.

A multicenter retrospective review was performed analyzing 1081 primary total hip replacements in 944 patients using the Harris Galante-I cementless acetabular component with screw fixation. All patients were followed up for a minimum of 5 years with a mean followup of 81 months. Linear polyethylene wear averaged 0.11 mm/year (range, 0-0.86 mm/year). Pelvic osteolysis was seen in 25 patients (2...

متن کامل

In vivo response to cross-linked polyethylene and polycarbonate-urethane particles.

This study was undertaken to examine macrophage response to polycarbonate-urethane, a proposed alternative material to polyethylene in acetabular components of total hip arthroplasty. Polyethylene wear debris from total joint replacements has been linked to osteolysis and implant lifespan. It has been shown in vitro, that polyethylene particles cleaned of endotoxin generate less of an inflammat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • The Journal of arthroplasty

دوره 30 9  شماره 

صفحات  -

تاریخ انتشار 2015