Optical quality differences between three multifocal intraocular lenses: bifocal low add, bifocal moderate add, and trifocal.
نویسندگان
چکیده
PURPOSE To compare the in vitro optical quality at different focal points of two new bifocal intraocular lenses (IOLs) and one new trifocal IOL. METHODS The AcrySof ReSTOR SV25T0 (+2.5 diopter [D] add) and the AcrySof ReSTOR SN6AD1 (+3.0 D add) with two main foci (Alcon Laboratories, Fort Worth, TX) and the AT LISA tri 839MP with three main foci (Carl Zeiss Meditec, Dublin, CA) were evaluated. The optical quality of the IOLs was measured with the PMTF optical bench (LAMBDA-X, Nivelles, Belgium). The optical quality of the IOLs was quantified by the modulation transfer function (MTF) at five different focal points (0.0, -1.5, -2.0, -2.5, and -3.0 D) and for 3.0- and 4.5-mm apertures. The through-focus MTF of the IOLs was also recorded. RESULTS For the 0.0 D focal point, the AcrySof ReSTOR (+2.5 D add) obtained the highest MTF values for all apertures. For the -2.5 D focal point, the AcrySof ReSTOR (+3.0 D add) showed the highest MTF values for 3.0 mm. For the -3.0 D focal point at 3.0- and 4.5-mm aperture, the best values were obtained with the AcrySof ReSTOR (+3.0 D add) and the AT LISA, respectively. For the -1.5 D focal point, the trifocal IOL provided better values. For the -2.0 D focal point, all IOLs provided comparable results. The through-focus MTF curves showed three and two peaks for the trifocal and bifocal IOLs, respectively. CONCLUSIONS The trifocal IOL provides a better optical quality at the -1.5 D focal point. However, the optical quality of the trifocal IOL significantly decreases compared to the bifocal IOLs at far distance and -2.5 D focal points.
منابع مشابه
In vitro optical quality measurements of three intraocular lens models having identical platform
BACKGROUND With recent advances in technology and introduction of new intraocular lens (IOL) models, surgeons today have the opportunity to choose from various optical designs, which can influence the postoperative quality of vision. In our laboratory study, we compared the optical quality of three different IOLs that use the identical platform and are produced by the same manufacturer. The stu...
متن کاملComparison of visual outcomes and subjective visual quality after bilateral implantation of a diffractive trifocal intraocular lens and blended implantation of apodized diffractive bifocal intraocular lenses
PURPOSE To compare the visual acuity (VA) and quality of vision between bilateral implantation of a trifocal intraocular lens (IOL) and blended bifocal IOLs with an intermediate add in the dominant eye and a near add in the nondominant eye. PATIENTS AND METHODS Patients with either trifocal or blended bifocal IOLs implanted were recruited after surgery. Subjects returned for a single diagnost...
متن کاملComparison of bifocal and trifocal diffractive and refractive intraocular lenses using an optical bench.
PURPOSE To assess the differences in optical performance of 9 multifocal IOLs using the same optical bench and to propose a possible comparison for surgeons. SETTING Rothschild Foundation, Paris, France. DESIGN Experimental study. METHODS Nine IOLs (Acrysof Restor +3.0 diopter [D] SN6AD1, Acrysof Restor +4.0 D SN60D3, Acrysof aspheric monofocal SN60WF, Acri.Lisa 366D, Finevision Micro F, ...
متن کاملLaboratory Evaluation of the Influence of Decentration and Pupil Size on the Optical Performance of a Monofocal, Bifocal, and Trifocal Intraocular Lens.
PURPOSE To assess the influence of decentration and aperture size on the optical quality of different intraocular lenses (IOLs) of the same material, body design, and refractive power using standardized optical bench testing. METHODS Using an optical bench set-up, an aspheric monofocal (CT ASPHINA 409M; Carl Zeiss Meditec, Jena, Germany), an aspheric diffractive bifocal (AT LISA 809M; Carl Ze...
متن کاملHalo and Through-Focus Performance of Four Diffractive Multifocal Intraocular Lenses.
PURPOSE To compare, as a function of pupil size, the through-focus performance and halo features of four diffractive intraocular lenses (IOLs). METHODS Three diffractive bifocal IOLs (ReSTOR +2.5 D SV25T0, Tecnis +2.75 D ZKB00, and AT LISA +3.75 D 809M) and a diffractive trifocal IOL (AT LISA tri +3.33 D, +1.66 D 839MP) were tested in vitro in a modified International Organization for Standar...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of refractive surgery
دوره 29 11 شماره
صفحات -
تاریخ انتشار 2013