Reproducibility of a peripheral quantitative computed tomography scan protocol to measure the material properties of the second metatarsal

نویسندگان

  • Elodie Chaplais
  • David Greene
  • Anita Hood
  • Scott Telfer
  • Verona du Toit
  • Davinder Singh-Grewal
  • Joshua Burns
  • Keith Rome
  • Daniel J Schiferl
  • Gordon J Hendry
چکیده

BACKGROUND Peripheral quantitative computed tomography (pQCT) is an established technology that allows for the measurement of the material properties of bone. Alterations to bone architecture are associated with an increased risk of fracture. Further pQCT research is necessary to identify regions of interest that are prone to fracture risk in people with chronic diseases. The second metatarsal is a common site for the development of insufficiency fractures, and as such the aim of this study was to assess the reproducibility of a novel scanning protocol of the second metatarsal using pQCT. METHODS Eleven embalmed cadaveric leg specimens were scanned six times; three times with and without repositioning. Each foot was positioned on a custom-designed acrylic foot plate to permit unimpeded scans of the region of interest. Sixty-six scans were obtained at 15% (distal) and 50% (mid shaft) of the second metatarsal. Voxel size and scan speed were reduced to 0.40 mm and 25 mm.sec(-1). The reference line was positioned at the most distal portion of the 2(nd) metatarsal. Repeated measurements of six key variables related to bone properties were subject to reproducibility testing. Data were log transformed and reproducibility of scans were assessed using intraclass correlation coefficients (ICC) and coefficients of variation (CV%). RESULTS Reproducibility of the measurements without repositioning were estimated as: trabecular area (ICC 0.95; CV% 2.4), trabecular density (ICC 0.98; CV% 3.0), Strength Strain Index (SSI) - distal (ICC 0.99; CV% 5.6), cortical area (ICC 1.0; CV% 1.5), cortical density (ICC 0.99; CV% 0.1), SSI - mid shaft (ICC 1.0; CV% 2.4). Reproducibility of the measurements after repositioning were estimated as: trabecular area (ICC 0.96; CV% 2.4), trabecular density (ICC 0.98; CV% 2.8), SSI - distal (ICC 1.0; CV% 3.5), cortical area (ICC 0.99; CV%2.4), cortical density (ICC 0.98; CV% 0.8), SSI - mid shaft (ICC 0.99; CV% 3.2). CONCLUSIONS The scanning protocol generated excellent reproducibility for key bone properties measured at the distal and mid-shaft regions of the 2(nd) metatarsal. This protocol extends the capabilities of pQCT to evaluate bone quality in people who may be at an increased risk of metatarsal insufficiency fractures.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Optimization of Radiation Dose in Cranial Computed Tomography among Adults: Assessment of Radiation Dose against Image Quality

Introduction: The rapid use of computed tomography (CT) scan is of great concern, due to increase in patients’ dose. Optimization of CT protocol is a vital issue in dose reduction. This study aimed to optimize radiation dose in cranial CT and assess modifications in image quality under radiation dose reduction. Material and Methods: A poly(me...

متن کامل

Strategy of computed tomography image optimization in cervical vertebra and neck soft tissue in emergency patients

Introduction: Due to the use of ionizing radiation in the CT scan, optimal parameters should be used to reduce the risk of cancer in patients who are constantly exposed to X-rays. The aim of this study was to optimize the parameters used in CT scan of cervical vertebrae and neck soft tissue with minimal loss of image quality   Materials and Methods: In this s...

متن کامل

Establishment of Diagnostic Reference Levels and Estimation of Effective Dose from Computed Tomography Head Scans at a Tertiary Hospital in South Africa

Introduction: Head scans are the most frequently performed computed tomography (CT) examinations worldwide. However, there is growing concern over the probability of increased cancer risks among the exposed populations. Diagnostic reference levels (DRLs) identify radiation dose that is not commensurate with clinical objectives. The aim of this study was to establish DR...

متن کامل

Fabrication of New 3D Phantom for the measurement of Geometric Distortion in Magnetic Resonance Imaging System

Introduction: Geometric distortion, an important parameter in neurology and oncology. The current study aimed to design and construct a new three-dimensional (3D) phantom using a 3D printer in order to measure geometric distortion and its 3D reproducibility. Material and Methods: In this study, a new phantom ...

متن کامل

Evaluation of Metal Artifact Reduction software in Computed Tomography

Introduction: The image quality of computed tomography (CT) can be seriously lowered by metal implants of patients. These implants are known to exert a significant impact on diagnostic accuracy due to artifacts. The current study aimed to assess the usefulness of Metal Artifact Reduction (MAR) software in the reduction of metal artifacts, in comparison to iterative rec...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 15  شماره 

صفحات  -

تاریخ انتشار 2014