Evaluation of a 2D diode array for IMRT quality assurance.

نویسندگان

  • Daniel Létourneau
  • Misbah Gulam
  • Di Yan
  • Mark Oldham
  • John W Wong
چکیده

BACKGROUND AND PURPOSE The QA of intensity modulated radiotherapy (IMRT) dosimetry is a laborious task. The goal of this work is to evaluate the dosimetric characteristics of a new 2D diode array (MapCheck from Sun Nuclear Corporation, Melbourne, Florida) and assess the role it can play in routine IMRT QA. MATERIAL AND METHODS Fundamental properties of the MapCheck such as reproducibility, linearity and temperature dependence are studied for high-energy photon beams. The accuracy of the correction for difference of diode sensitivity is also assessed. The diode array is benchmarked against film and ion chambers for conventional and IMRT treatments. The MapCheck sensitivity to multileaf collimator position errors is determined. RESULTS The diode array response is linear with dose up to 295 cGy. All diodes are calibrated to within +/-1% of each other, and mostly within +/-0.5%. The MapCheck readings are reproducible to within a maximum SD of +/-0.15%. A temperature dependence of 0.57%/ degrees C was noted and should be taken into account for absolute dosimetric measurement. Clinical performance of the MapCheck for relative and absolute dosimetry is demonstrated with seven beam (6 MV) head and neck IMRT plans, and compares well with film and ion chamber measurements. Comparison to calculated dose maps demonstrates that the planning system model underestimates the dose gradients in the penumbra region. CONCLUSIONS The MapCheck offers the dosimetric characteristics required for performing both relative and absolute dose measurements. Its use in the clinic can simplify and reduce the IMRT QA workload.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

3D Analysis of Intensity-Modulated Radiation Therapy Quality Assurance Measurement using a 2D Diode Array

Intensity-modulated radiation therapy (IMRT) quality assurance (QA) is often performed using a 2D device and compares measured and computed fluence maps to determine if a field passes or fails certain dose and position criteria. The effects of a measured deviation to the 3D patient spatial dosimetry and dose volume histogram (DVH) are largely unknown because they cannot be analyzed using commer...

متن کامل

Evaluation of a fast method of EPID-based dosimetry for IMRT and Comparison with 3D EPID-based dosimetry system using conventional two- and three-dimensional detectors for VMAT

Introduction: Electronic portal imaging devices (EPIDs) could potentially be useful for intensity-modulated radiation therapy (IMRT) and VMAT QA. The data density, high resolution, large active area, and efficiency of the MV EPID make it an attractive option. However, EPIDs were designed as imaging devices, not dosimeters, and as a result they do not inherently measure dose in ...

متن کامل

A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.

An analysis is made of a two-dimensional array of diodes that can be used for measuring dose generated in a plane by a radiation beam. This measuring device is the MapCHECK Model 1175 (Sun Nuclear, Melbourne, FL). This device has 445 N-type diodes in a 22 x 22 cm2 2-D array with variable spacing. The entire array of diodes is easily calibrated to allow for measurements in absolute dose. For IMR...

متن کامل

A critical evaluation of the PTW 2D‐ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification

Quality assurance (QA) for intensity- and volumetric-modulated radiotherapy (IMRT and VMAT) has evolved substantially. In recent years, various commercial 2D and 3D ionization chamber or diode detector arrays have become available, allowing for absolute verification with near real time results, allowing for streamlined QA. However, detector arrays are limited by their resolution, giving rise to...

متن کامل

2D linear array device as a quality assurance tool in brachytherapy applications

Background: External beam radiotherapy and brachytherapy plays a vital role in the management of cancer cervix.  High dose rate brachytherapy is being presently used worldwide for the brachytherapy applications. At present, 2-Dimensional linear array detectors   are the most common QA tool used for pretreatment patient specific quality assurance in external beam radiotherapy alon...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology

دوره 70 2  شماره 

صفحات  -

تاریخ انتشار 2004