Wavelet Domain Radiofrequency Pulse Design Applied to Magnetic Resonance Imaging
نویسندگان
چکیده
A new method for designing radiofrequency (RF) pulses with numerical optimization in the wavelet domain is presented. Numerical optimization may yield solutions that might otherwise have not been discovered with analytic techniques alone. Further, processing in the wavelet domain reduces the number of unknowns through compression properties inherent in wavelet transforms, providing a more tractable optimization problem. This algorithm is demonstrated with simultaneous multi-slice (SMS) spin echo refocusing pulses because reduced peak RF power is necessary for SMS diffusion imaging with high acceleration factors. An iterative, nonlinear, constrained numerical minimization algorithm was developed to generate an optimized RF pulse waveform. Wavelet domain coefficients were modulated while iteratively running a Bloch equation simulator to generate the intermediate slice profile of the net magnetization. The algorithm minimizes the L2-norm of the slice profile with additional terms to penalize rejection band ripple and maximize the net transverse magnetization across each slice. Simulations and human brain imaging were used to demonstrate a new RF pulse design that yields an optimized slice profile and reduced peak energy deposition when applied to a multiband single-shot echo planar diffusion acquisition. This method may be used to optimize factors such as magnitude and phase spectral profiles and peak RF pulse power for multiband simultaneous multi-slice (SMS) acquisitions. Wavelet-based RF pulse optimization provides a useful design method to achieve a pulse waveform with beneficial amplitude reduction while preserving appropriate magnetization response for magnetic resonance imaging.
منابع مشابه
Designing and Fabrication of a New Radiofrequency Planar microcoil for mini-Nuclear Magnetic Resonance
Introduction Radiofrequency planar microcoils are used to increase the resolution of magnetic resonance images of small samples. In this study, we aimed to design and fabricate a spiral planar microcoil constructed on a double-sided printed circuit board (PCB). It has four rings with an internal diameter of 241 microns tuned and matched at 63.8 MHz. Materials and Methods To achieve the maximum ...
متن کاملApproximate Fourier domain expression for Bloch-Siegert shift.
PURPOSE In this study, a new simple Fourier domain-based analytical expression for the Bloch-Siegert (BS) shift-based B1 mapping method is proposed to obtain |B1+| more accurately while using short BS pulse durations and small off-resonance frequencies. THEORY AND METHODS A new simple analytical expression for the BS shift is derived by simplifying the Bloch equations. In this expression, the...
متن کاملMR image encoding by spatially selective rf excitation: An analysis using linear response models
Spatially selective radiofrequency (rf) excitation has been applied to encode magnetic resonance (MR) images using non-Fourier basis sets such as wavelets. While non-Fourier–encoded MRI has the advantage of flexibility allowing for various adaptive imaging strategies, pulse sequence implementation for spatially selective rf excitation presents unique problems. Two linear response models are des...
متن کاملOptimal control of coupled spin dynamics: design of NMR pulse sequences by gradient ascent algorithms.
In this paper, we introduce optimal control algorithm for the design of pulse sequences in NMR spectroscopy. This methodology is used for designing pulse sequences that maximize the coherence transfer between coupled spins in a given specified time, minimize the relaxation effects in a given coherence transfer step or minimize the time required to produce a given unitary propagator, as desired....
متن کاملFour dimensional spectral-spatial fat saturation pulse design.
PURPOSE The conventional spectrally selective fat saturation pulse may perform poorly with inhomogeneous amplitude of static (polarizing) field (B0 ) and/or amplitude of (excitation) radiofrequency field (B1 ) fields. We propose a four dimensional spectral-spatial fat saturation pulse that is more robust to B0/B1 inhomogeneity and also shorter than the conventional fat saturation pulse. THEOR...