Two-Dimensional Capacitive Micromachined Ultrasonic Transducer (CMUT) Arrays for a Miniature Integrated Volumetric Ultrasonic Imaging System

نویسندگان

  • X. Zhuang
  • I. O. Wygant
  • D. T. Yeh
  • A. Nikoozadeh
  • O. Oralkan
  • A. S. Ergun
  • B. T. Khuri-Yakub
چکیده

We have designed, fabricated, and characterized two-dimensional 16x16-element capacitive micromachined ultrasonic transducer (CMUT) arrays. The CMUT array elements have a 250-μm pitch, and when tested in immersion, have a 5-MHz center frequency and 99% fractional bandwidth. The fabrication process is based on standard silicon micromachining techniques and therefore has the advantages of high yield, low cost, and ease of integration. The transducers have a Si3N4 membrane and are fabricated on a 400-μm thick silicon substrate. A low parasitic capacitance through-wafer via connects each CMUT element to a flip-chip bond pad on the back side of the wafer. Each through-wafer via is 20 μm in diameter and 400 μm deep. The interconnects form metal-insulator-semiconductor (MIS) junctions with the surrounding high-resistivity silicon substrate to establish isolation and to reduce parasitic capacitance. Each through-wafer via has less than 0.06 pF of parasitic capacitance. We have investigated a Au-In flip-chip bonding process to connect the 2D CMUT array to a custom integrated circuit (IC) with transmit and receive electronics. To develop this process, we fabricated fanout structures on silicon, and flip-chip bonded these test dies to a flat surface coated with gold. The average series resistance per bump is about 3 Ohms, and 100% yield is obtained for a total of 30 bumps.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Volumetric Imaging Using 2D Capacitive Micromachined Ultrasonic Transducer Arrays (CMUTs): Initial Results

This paper presents the first volumetric images obtained using a 2D CMUT array with throughwafer via interconnects. An 8×16-element portion of a 32×64-element array flip-chip bonded onto a glass fanout chip was used in the experiments. This study experimentally demonstrates that 2D CMUT arrays can be fabricated with high yield using silicon micromachining processes, individual electrical connec...

متن کامل

Medical imaging using capacitive micromachined ultrasonic transducer arrays.

We are investigating the use of capacitive micromachined ultrasonic transducers (cMUT's) for use in medical imaging. We propose an ultrasound probe architecture designed to provide volumetric ultrasound imaging from within an endoscope channel. A complete automated experimental system has been implemented for testing the imaging performance of cMUT arrays. This PC-based system includes custom-d...

متن کامل

Fabrication and Characterization of 1-Dimensional and 2-Dimensional Capacitive Micromachined Ultrasonic Transducer (CMUT) Arrays for 2- Dimensional and Volumetric Ultrasonic Imaging

AbsWmI-Capacitive hlirromarhined tiltrasonic Transducers (Ch1l:Ts) were introduced about a decade ago as an alternate method of pnerating and detecting ultrasound. Since their introduction, considerable research hss been done lo characterize CNUTs. They have been shown to have broad frequency bandwidth and very good sensitivity. Besides, CMtlTs are built on silicon using standard surface microm...

متن کامل

Capacitive Micromachined Ultrasonic Transducer Technology for Medical Ultrasound Imaging

Capacitive micromachined ultrasonic transducer (cMUT) technology has been recognized as an attractive alternative to the more traditional piezoelectric transducer technology in medical ultrasound imaging for several years now. There are mainly two reasons for the interest in this technology: Micromachining is derived from the integrated circuit technology and therefore shares the well-known adv...

متن کامل

Capacitive micromachined ultrasonic transducer (CMUT) arrays for medical imaging

Capacitive micromachined ultrasonic transducers (CMUTs) bring the fabrication technology of standard integrated circuits into the field of ultrasound medical imaging. This unique property, combined with the inherent advantages of CMUTs in terms of increased bandwidth and suitability for new imaging modalities and high frequency applications, have indicated these devices as new generation arrays...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2005