Effects of ECG gating and postprocessing techniques on 3D MDCT of the bronchial tree.
نویسندگان
چکیده
OBJECTIVE Our goal was to determine the impact of ECG gating and different postprocessing techniques on 3D imaging of the bronchial tree. SUBJECTS AND METHODS. Retrospective ECG-gated MDCT and non-ECG-gated MDCT of the chest were performed in 25 patients. ECG-gated MDCT data were reconstructed mid diastole using a fixed interval of -400 msec in 25 patients and then additionally at -200, -300, and -500 msec in 10 of those patients. Shaded surface display and volume rendering of the bronchial tree combined with virtual bronchoscopy were performed using all data sets. The extent of bronchial tree visualization in shaded surface display-virtual bronchoscopy and volume rendering-virtual bronchoscopy and the presence of artifacts in volume-rendered images were scored by three blinded reviewers. The effective radiation doses of the ECG-gated and nongated acquisitions were compared. RESULTS The summary scores of all bronchial segments for gated shaded surface display-virtual bronchoscopy and gated volume rendering-virtual bronchoscopy did not differ significantly. The summary scores for nongated shaded surface display-virtual bronchoscopy and nongated volume rendering-virtual bronchoscopy were not significantly different. Non-gated acquisition yielded significantly better visualization of the bronchial tree for both post-processing techniques, regardless of the time interval used for reconstruction of the ECG-gated series. Artifact scores in volume-rendered images were significantly higher for ECG-gated MDCT compared with nongated MDCT. Effective radiation dose was significantly higher for the ECG-gated acquisition. CONCLUSION Given the advantage of volume rendering for representing the entire data set and given the lower radiation dose and better 3D image quality of nongated acquisition, volume rendering performed on nongated MDCT data is the method of choice for 3D visualization of the bronchial tree.
منابع مشابه
Retrospectively ECG-gated multi-detector row CT of the chest: does ECG-gating improve three-dimensional visualization of the bronchial tree?
PURPOSE To determine the impact of retrospectively ECG-gated multi-detector row CT (MDCT) on three-dimensional (3D) visualization of the bronchial tree and virtual bronchoscopy (VB) as compared to non-ECG-gated data acquisition. MATERIALS AND METHODS Contrast-enhanced retrospectively ECG-gated and non-ECG-gated MDCT of the chest was performed in 25 consecutive patients referred for assessment...
متن کاملNon-coronary findings in cardiac CT
Gated multi-detector CT (MDCT) is now recognised as an excellent and robust application for non-invasive assessment of coronary arteries, and in many centres is routinely used. Some studies have shown that MDCT scanners with 64-detector row have a 99% negative predictive value for the detection of coronary artery stenosis. MDCT images obtained with ECG gating are superior to the standard (non-E...
متن کاملLow kilovolt “prospective ECG-triggering” vs. “retrospective ECG-gating” coronary CTA: comparison of image quality and radiation dose
Background: To compare image quality and radiation doses of low kilovolt (kV) “prospective ECG-triggering” (PT) and standard “retrospective ECG-gating” (RG) coronary computed tomography (CT) angiography. Materials and Methods: A total of 101 consecutive patients (76 males, 25 females; mean age: 55.44 ± 8.28 years) with low-to-intermediate risk status for coronary artery disease and with a body ...
متن کاملComparison of non-gated vs. electrocardiogram-gated 64-detector-row computed tomography for integrated electroanatomic mapping in patients undergoing pulmonary vein isolation.
AIMS To compare non-gated vs. electrocardiogram (ECG)-gated 64-detector-row computed tomography (MDCT) of the left atrium (LA) for integrated electroanatomic mapping (EAM) in patients with paroxysmal atrial fibrillation (AF). METHODS AND RESULTS Twenty-nine consecutive patients with paroxysmal AF underwent MDCT prior to pulmonary vein isolation (PVI). All patients were in sinus rhythm both du...
متن کاملRadiation dose reduction and coronary assessability of prospective electrocardiogram-gated computed tomography coronary angiography: comparison with retrospective electrocardiogram-gated helical scan.
OBJECTIVES The aim of this study was to evaluate radiation dose and coronary assessability of a prospective electrocardiogram (ECG)-gated scan by 64-slice multidetector (row) computed tomography (MDCT)-coronary angiography (CA) compared with a retrospective ECG-gated helical scan. BACKGROUND The 64-slice MDCT-CA has been widely used; however, a high radiation dose by 64-slice MDCT-CA has been...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- AJR. American journal of roentgenology
دوره 183 1 شماره
صفحات -
تاریخ انتشار 2004