A new approach to measure dwell position inaccuracy in HDR ring applicators – quantification and corrective QA
نویسندگان
چکیده
As part of quality assurance (QA) in high dose rate brachytherapy, it is necessary to verify that the source dwell positions correspond to the radiographic markers used in simulation and treatment planning. The procedure is well established for linear tandem applicators. However, with the advent of ring applicators, this has become more critical and challenging. This work describes a new approach to determine positional inaccuracies for ring applicators in which the dummy markers are imaged just once and their dwell positions characterized with respect to an applicator-defined axis. The radiograph serves as a reference for dummy markers for comparison with all subsequent measurements in which the active sources are autoradiographed at different offsets - thus obviating the back-and-forth transferring of setup between afterloader and simulator. The method has been used specifically to characterize the Varian GammaMed 60° ring applicator, but it may be adapted to any other applicator. The results show that an offset of 1-2 mm minimizes the overall inaccuracy to within ± 2 mm.
منابع مشابه
SU-E-T-138: Quantification of Dwell Position Inaccuracy in Varian GammaMed HDR Titanium Ring Applicators.
PURPOSE To quantify the dwell position inaccuracy in Titanium ring applicators and develop a test to be performed quarterly, after source exchange. METHODS All three rings from our Titanium kit (30, 45, 60 deg.) were used for this study. EDR2 film was placed on the Simulator table and a ring was taped to the film, with a solid water slab as buildup. A 1-cm spacing dummy wire was inserted into...
متن کاملDwell position inaccuracy in the Varian GammaMed HDR ring applicator
Varian has issued two Product Notification Letters warning of known inaccuracies in dwell positions for their GammaMed HDR ring applicator. This inaccuracy was measured for two sets of applicators. Autoexposed radiographs were taken of the HDR source at different dwell positions and analyzed per Varian recommendations using tools within the BrachyVision treatment planning program. Comparison be...
متن کاملQuality assurance measurements for high‐dose‐rate brachytherapy without film
The purpose of this study was to develop new and modified tools that allow HDR brachytherapy quality assurance tests to be carried out efficiently without film, video cameras, stopwatches, and mechanical rulers; and to devise methods that use these new tools for daily and quarterly check procedures, which are efficient and provide increased accuracy compared to previous methods. The HDR brachyt...
متن کاملCommissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets
Studies have shown that source dwells within Varian's HDR CT/MR compatible ring applicators can deviate from intended positions by several millimeters. Quantifying this offset is an important part of commissioning. The aims of this study were to: 1) determine the reproducibility of the offset, 2) study the interobserver variation in the offset's measurement, and 3) quantify the dosimetric impac...
متن کاملIndependent assessment of source transit time for the BEBIG SagiNova® high dose rate brachytherapy afterloader
Introduction: The dwell time and transit time components contribute to the overall delivered dose to patients in high dose rate (HDR) brachytherapy treatments. The transit time results from source entry and exit as well as source movements between dwell positions. It depends on various parameters such as the source speed profile, source indexer distance to dwell position, and ...