Oxygen contamination of the nitrous oxide pipeline supply.
نویسندگان
چکیده
Midogas (CIG Australia) nitrous oxide/oxygen blenders are commonly used in delivery wards in Australia. In this case report we describe an incident where a small hole in the diaphragm of the oxygen failure alarm in a Midogas nitrous oxide/oxygen blender led to retrograde flow of oxygen contaminating the nitrous oxide pipeline supply to the operating theatres and the delivery ward. This caused a reduced level of nitrous oxide to the patient in theatre, but there was no adverse outcome. However, if the oxygen pipeline pressure had been lower than the nitrous oxide pipeline pressure, the contamination would have been reversed, potentially resulting in a hypoxic mixture being delivered to many patients. Suggestions are made as to a method of prevention of this problem.
منابع مشابه
Contamination of anaesthetic gases with nitric oxide and its influence on oxygenation: study in patients undergoing open heart surgery.
BACKGROUND Nitric oxide is important in vasomotor regulation. Contamination of anaesthetic gases with nitric oxide could affect gas exchange. METHODS We measured oxygenation and nitric oxide concentrations in the inspiratory and expiratory limb of the ventilator circuit in patients about to have cardiac surgery. Measurements were made before surgery when the circulation and respiratory condit...
متن کاملAn assessment of operating room environment air contamination with nitrous oxide and halothane and some scavenging methods.
This study was designed to assess operating room contamination with nitrous oxide and halothane when nitrous oxide 3 1/min and oxygen 2 1/min containing halothane one per cent were passed into a semiclosed circle absorber system from which the patients' lungs were ventilated with an Ohio ventimeter through a cuffed tracheal tube, with the exhaled gas vented to the floor. The normal room ventila...
متن کاملCardiac arrhythmia during outpatient dental anaesthesia: the advantages of a controlled ventilation technique.
Two techniques for endotracheal anaesthesia in dental outpatients have been compared: a spontaneous ventilation technique using nitrous oxide, oxygen and halothane, and a controlled ventilation technique using nitrous oxide and oxygen, with tubocurarine and fentanyl. The latter technique has several advantages: abolition of arrhythmias induced by surgery, more rapid and hence safer recovery, av...
متن کاملTwo cases of poisoning by contamination of nitrous oxide with higher oxides of nitrogen during anaesthesia.
It will now be widely known that contamination with a higher oxide of nitrogen occurred in a batch of nitrous oxide cylinders which were filled by the British Oxygen Company near the beginning of September 1966. When one considers the vast numbers of cylinders of nitrous oxide diat have been used during the last 40 years, contamination is obviously extremely rare; but, despite the care taken by...
متن کاملمقایسه روش های اینداکشن استنشاقی Single breath inhalation و Tidal volume technique با اینداکشن وریدی (نسدونال، ساکسینیل)، دانشگاه علوم پزشکی تهران، 1379
The induction characteristic of halothane in nitrous oxide and oxygen were compared with halothane in oxygen alone and intravenous induction with thiopentum and succinylcholine.Materials and Methods: A vital capacity technique was used for the gaseous induction groups using a mapleson F system and a 1-litre reservoir bag. four end points of anesthesia were recorded: time to cessation of hand gr...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Anaesthesia and intensive care
دوره 26 5 شماره
صفحات -
تاریخ انتشار 1998