16-Channel biphasic current-mode programmable charge balanced neural stimulation
نویسندگان
چکیده
BACKGROUND Neural stimulation is an important method used to activate or inhibit action potentials of the neuronal anatomical targets found in the brain, central nerve and peripheral nerve. The neural stimulator system produces biphasic pulses that deliver balanced charge into tissue from single or multichannel electrodes. The timing and amplitude of these biphasic pulses are precisely controlled by the neural stimulator software or imbedded algorithms. Amplitude mismatch between the anodic current and cathodic current of the biphasic pulse will cause permanently damage for the neural tissues. The main goal of our circuit and layout design is to implement a 16-channel biphasic current mode programmable neural stimulator with calibration to minimize the current mismatch caused by inherent complementary metal oxide semiconductor (CMOS) manufacturing processes. METHODS This paper presents a 16-channel constant current mode neural stimulator chip. Each channel consists of a 7-bit controllable current DAC used as sink and source current driver. To reduce the LSB quantization error and the current mismatch, an automatic calibration circuit and flow diagram is presented in this paper. There are two modes of operation of the stimulator chip-namely, stimulation mode and calibration mode. The chip also includes a digital interface used to control the stimulator parameters and calibration levels specific for each individual channel. RESULTS This stimulator Application Specific Integrated Circuit (ASIC) is designed and fabricated in a 0.18 μm High-Voltage CMOS technology that allows for ±20 V power supply. The full-scale stimulation current was designed to be at 1 mA per channel. The output current was shown to be constant throughout the timing cycles over a wide range of electrode load impedances. The calibration circuit was also designed to reduce the effect of CMOS process variation of the P-channel metal oxide semiconductor (PMOS) and N-channel metal oxide semiconductor (NMOS) devices that will result in charge delivery to have less than 0.13% error. CONCLUSIONS A 16-channel integrated biphasic neural stimulator chip with calibration is presented in this paper. The stimulator circuit design was simulated and the chip layout was completed. The chip layout was verified using design rules check (DRC) and layout versus schematic (LVS) design check using computer aided design (CAD) software. The test results we presented show constant current stimulation with charge balance error within 0.13% least-significant-bit (LSB). This LSB error was consistent throughout a variety stimulation patterns and electrode load impedances.
منابع مشابه
Asymmetric Rectangular Waveform in Stimulation with High Frequency Alternating Current Reduces the Threshold for Neural Conduction Block
Introduction Abnormal neural impulses in the nervous system may lead to various diseases and disabilities. High frequency alternating currents (HFAC) has been used to block the propagation of such impulses and improve the symptoms or disabilities. The technique is safe, reversible, and relatively selective, and its reliability, the optimum stimulation parameters, and elimination of the onset re...
متن کاملA Programmable Neural Interface for Investigating Arbitrary Stimulation Strategies
This paper presents a microcontroller-based neural interface capable of producing arbitrary current-mode waveforms for use in developing and evaluating electrical stimulation strategies. The presented application-specific customisation focuses on charge-balanced biphasic waveforms delivered to a bipolar electrode configuration, although the hardware is capable of interfacing with other electrod...
متن کاملOffset prediction for charge-balanced stimulus waveforms.
Functional electrical stimulation with cuff electrodes involves the controlled injection of current into an electrically excitable tissue for sensory or motor rehabilitation. Some charge injected during stimulation is 'lost' at the electrode-electrolyte interface when the charge carrier is translated from an electron to an ion in the solution. The process of charge injection through chemical re...
متن کاملElectrical stimulation of the auditory nerve: II. Effect of stimulus waveshape on single fibre response properties.
To investigate the generation of action potentials by electrical stimulation we studied the response of auditory nerve fibres (ANFs) to a variety of stimulus waveforms. Current pulses were presented to longitudinal bipolar scala tympani electrodes implanted in normal and deafened cochleae. Capacitively coupled monophasic current pulses evoked single ANF responses that were more sensitive to one...
متن کاملA 0.042 mm^2 programmable biphasic stimulator for cochlear implants suitable for a large number of channels
This paper presents a compact programmable biphasic stimulator for cochlear implants. By employing double-loop negative feedback, the output impedance of the current generator is increased, while maximizing the voltage compliance of the output transistor. To make the stimulator circuit compact, the stimulation current is set by scaling a reference current using a two stage binary-weighted trans...
متن کامل