Optimization and Validation of a High-performance Liquid Chromatographic Method for Determination of Α- and Γ-tocopherol in Rat Plasma and Erythrocytes
نویسندگان
چکیده
An HPLC method for determination of αand γ-tocopherol in rat blood has been optimized and validated. The optimum means of separation and analysis of tocopherol homologs were selected on the basis of procedures found in the literature; documentary evidence is presented which shows precision and accuracy were satisfactory. Spectrofluorimetric detection was performed after separation on a Separon SGX NH2 analytical column with n-hexane–2-propanol, 97:3, as mobile phase; racemic tocol was used as internal standard. A suspension of erythrocytes was saponified and tocopherols were extracted in the presence of BHT, ascorbic acid, and pyrogallol. Plasma samples were extracted without saponification but with antioxidant protection. Validation revealed that intra-day and inter-day RSD in plasma were 0.78% and 4.15%, respectively, for α-tocopherol and 4.03% and 13.155%, respectively, for γ-tocopherol. Intra-day RSD for erythrocytes were 3.75% and 2. 89% for α-tocopherol and γ-tocopherol, respectively. Absolute percentage recovery from plasma and erythrocytes was 99.8 ± 7.5 and 94 ± 9.0, respectively, for α-tocopherol and 100.2 ± 7.5 and 96.9 ± 7.8, respectively, for γ-tocopherol. Limits of detection in plasma and erythrocytes were 0.005 and 0.002 μg mL, respectively, for α-tocopherol and 0.002 and 0.0015 μg mL, respectively, for γ-tocopherol. Limits of quantification in plasma and erythrocytes were 0.015 and 0.006 μg mL, respectively, for α-tocopherol and 0.006 and 0.0045 μg mL, respectively, for γ-tocopherol.
منابع مشابه
Simple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma
A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-...
متن کاملSimple High-Performance Liquid Chromatographic Method for Determination of Ciprofloxacin in Human Plasma
A rapid, simple and sensitive high-performance liquid chromatography method was developed for determination of ciprofloxacin in plasma by means of ultraviolet detection. Ofloxacin was used as an internal standard and separation carried on a Novapak C18 column using a mobile phase of 0.01 M phosphate buffer (pH =2.6): methanol (82: 18 v/v). Extraction of drug was performed from plasma by liquid-...
متن کاملComparison of Microbiological and High-Performance Liquid Chromatographic Methods for Determination of Clarithromycin Levels in Plasma
An agar well diffusion bioassay method for determination of clarithromycin in human plasma, using Micrococcus Luteus ATCC 9341 as the assay organism, was compared with a selective high-performance liquid chromatographic (HPLC) method with UV detection. Spiked plasma was used to prepare standard and control samples for both methods. The results of the bioassay analyses with spiked plasma samples...
متن کاملComparison of Microbiological and High-Performance Liquid Chromatographic Methods for Determination of Clarithromycin Levels in Plasma
An agar well diffusion bioassay method for determination of clarithromycin in human plasma, using Micrococcus Luteus ATCC 9341 as the assay organism, was compared with a selective high-performance liquid chromatographic (HPLC) method with UV detection. Spiked plasma was used to prepare standard and control samples for both methods. The results of the bioassay analyses with spiked plasma samples...
متن کاملRapid High Performance Liquid Chromatographic Determination of Risperidone in Human Plasma
A simple, rapid and sensitive high-performance liquid chromatographic (HPLC) method for the determination of risperidone in human plasma was developed. An HPLC system based on a Nucleosil C8 column (150?4 mm) and a UV detector (?= 280 nm) were used. A mixture of sodium dihydrogen phosphate buffer-acetonitrile (55:45, v/v) adjusted to pH 6.0 at a flow rate of 1.5 ml min-1 was used as mobile phas...
متن کامل