Evaluation of dosimetric effect of leaf position in a radiation field of an 80‐leaf multileaf collimator fitted to the LINAC head as a tertiary collimator
نویسندگان
چکیده
This study evaluates changes in the dosimetric characteristics of a Varian Millennium 80-leaf multileaf collimator (MLC) in a radiation field. In this study, dose rate, scatter factor, percentage depth dose, surface dose and dose in the buildup region, beam profile, flatness and symmetry, and penumbra width measurements were made for 6-MV and 15-MV photon beams. Analysis of widths between 50% dose levels of the beam profiles to reflect the field size at the level of profile measurement shows a significant difference between the fields defined by MLC and/or jaws and MLC (zero gap) and the fields defined by jaws only. The position of the MLC leaves in the radiation field also significantly affects scatter factors. A new relationship has, therefore, been established between the scatter factors and the position of the MLC, which will indeed be useful in the dose calculation for irregular fields. Penumbra widths increase with field size and were higher for fields defined by jaws and/or MLC than jaws and MLC (zero gap) by 1.5 mm to 4.2 mm and 3.8 mm to 5.0 mm, for 6-MV, and 1.5 mm to 2.4 mm and 3.0 mm to 5.6 mm, for 15-MV, at 20% to 80% and 10% to 90% levels, respectively. The surface dose and the dose in the buildup region were smaller for fields defined by jaws and MLC (zero gap) than the fields defined by jaws and/or MLC for both photon energies. No significant differences were found in percentage depth dose beyond dmax, beam profiles above 80% dose level, and flatness and symmetry for both energies. The results of this study suggest that while one collects linear accelerator beam data with a MLC, the effects of the positions of the MLC leaves play an important role in dosimetric characteristics of 3D conformal radiation therapy as well as intensity-modulated radiotherapy.
منابع مشابه
Physical and Dosimetric Aspect of Euromechanics Add-on Multileaf Collimator on Varian Clinac 2100 C/D
Background:Â Before treatment planning and dose delivery, quality assurance of multi-leaf collimator (MLC) has an important role in intensity-modulated radiation therapy (IMRT) due to the creation of multiple segments from optimization process.Objective: The purpose of this study is to assess the quality control of MLC leaves using EBT3 Gafchromic films.Material and Methods: Leaf Position accur...
متن کاملThe Dosimetric Effects of Different Multileaf Collimator Widths on Physical Dose Distributions
Introduction: Geometric changes in the multileaf collimator (MLC) led to dosimetric considerations in intensity-modulated radiation therapy (IMRT) due to the number and size of the pixels in the intensity map, which are determined by the MLC leaf width. In this study, we evaluated the dosimetric effects of different MLC widths on physical dose distributions for IMRT plans. Materials and Method...
متن کاملBeam characteristic of P.E collimators Add-on multileaf collimator
Introduction: Multileaf collimators (MLCs) are computer-controlled devices which are Useful and cost-effective tools for conformal therapy and intensity modulation radiotherapy (IMRT). Nowadays MLCs are important and standard configuration of the new medical linear accelerators which are substitute of conventional Blocks and has eliminated the ...
متن کاملDosimetric and qualitative analysis of kinetic properties of millennium 80 multileaf collimator system for dynamic intensity modulated radiotherapy treatments.
The aim of this paper is to analyze the positional accuracy, kinetic properties of the dynamic multileaf collimator (MLC) and dosimetric evaluation of fractional dose delivery for the intensity modulated radiotherapy (IMRT) for step and shoot and sliding window (dynamic) techniques of Varian multileaf collimator millennium 80. Various quality assurance tests such as accuracy in leaf positioning...
متن کاملDesign and Fabrication of the Mechanical Part of a Prototype Multileaf Collimator System
Introduction: Multileaf collimator (MLC) is state of the art among the radiation field shaping systems for conformal radiotherapy techniques. Some advantages of this system are as follow: reduction of treatment time, higher precision in conforming the radiation field to the projection of the target volume and lowering the chance of errors made by technologists. The purpose of this research was ...