Spontaneous ventilation with the Bain anaesthetic system.
نویسندگان
چکیده
Measurements of ventilation and of inspired gas composition were made while volunteers breathed a non-anaesthetic gas through a Bain anaesthetic system. It was found that rebreathing occurred when the fresh gas flow was between two-and-a-half and three times the minute volume. Fresh gas flows at least three times the minute volume appear to be necessary to prevent rebreathing when using this system.
منابع مشابه
The oxygen and volatile anaesthetic delivery characteristics during spontaneous ventilation using a non-rebreathing coaxial anaesthesia system.
The gas delivery characteristics during simulated spontaneous ventilation of a non-rebreathing adaptation of the Bain circuit were examined. Tidal volumes were varied from 200 to 600 ml and respiratory rate from 8 to 20 breaths per minute obtaining characteristics for a minute volume range from 2.4 to 12 litres/minute. The minimum inspired oxygen concentration with 1 litre/minute 100% oxygen fr...
متن کاملWill we ever have a universal anaesthetic breathing system.
In this issue of the JOURNAL, Humphrey and his colleagues describe a version of the A.D.E. anaesthetic breathing system 1'2 which they feel is optimal for all patients. Humphrey et al . have shown that the "A.D.E. low-flow universal" system functions as a modified D (Mapleson's classification 3) or Bain circuit for controlled ventilation and, with the switch of a single lever, as a modified Map...
متن کاملEnclosed afferent reservoir breathing systems. Description and clinical evaluation.
A new group of breathing systems, namely the Enclosed Afferent Reservoir (EAR) systems, is described. They allow for the selective elimination of alveolar gas in association with both spontaneous and controlled ventilation. A comparison with the Bain system in controlled ventilation demonstrates greater efficiency in eliminating carbon dioxide. A fresh gas flow (VF) of 70 ml kg-1 min-1 using an...
متن کاملHow rebreathing anaesthetic systems control PaCO2: studies with a mechanical and a mathematical model.
Results from a proposed equation for rebreathing systems, (formula: see text), were compared with results from a mechanical model "lung" ventilated either with a Bain Breathing Circuit, or a circle system (Eger-Ethans type A) without soda lime. When values for carbon dioxide excretion (VCO2), dead spacetidal volume ratio (VD/VT), minute volume ventilation (VE), and fresh gas flow (VF) were vari...
متن کاملSingle lever Hum - phrey A . D . E . low flow universal anaesthetic breathing system Part I : Comparison with dual lever A . D . E . ,
The single lever Humphrey A.D.E. anaesthetic system, in both coaxial and parallel (non-coaxial)forms, has recently been introduced. In principle the system offers efficient "universal" function by combining the advantages of Mapleson A, D and E systems. A within-patient comparison of its function in the Mapteson A mode (lever up) in spontaneously-breathing anaesthetized subjects was made to tha...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- British journal of anaesthesia
دوره 49 12 شماره
صفحات -
تاریخ انتشار 1977