Microbubble transmission during cardiotomy infusion of a hardshell venous reservoir with integrated cardiotomy versus a softshell venous reservoir with separated cardiotomy: an in vitro comparison.
نویسندگان
چکیده
An important mechanism for postoperative cognitive impairment after cardiac surgery using cardiopulmonary bypass (CPB) is microemboli. One component of the CPB circuit-the cardiotomy-is a major source of gaseous microemboli because it aspirates significant volumes of air with blood from the operative field and intracardiac chambers. Cardiotomies are either integrated within an open hardshell venous reservoir (IC-HSVR) or are a separate canister attached to a softshell collapsible venous reservoir bag (SC-SSVR). The purpose of this study was to compare the Medtronic IC-HSVR (Affinity NT CVR) with Medtronic's SC-SSVR (CB 1351, CBMVR 1600) in terms of relative microbubble transmission during cardiotomy infusion. A recirculating in vitro circuit primed with blood was used to compare the two cardiotomy-reservoir systems with the venous reservoir in the SC-SSVR further assessed in a fully closed or partially open state (SC-SSVR-closed; SC-SSVR-open). Microbubbles were detected using a GAMPT BC100 Doppler system in the outflow line of the venous reservoir. Measurements were taken before (baseline) and after aerated prime was pumped into the cardiotomy while altering pump flow rates (3 L/min; 5 L/min) and reservoir prime volumes (400 mL; 900 mL). Infusing cardiotomy blood into the venous reservoir was associated with an increase in microbubbles and bubble volume transmitted by both cardiotomy-reservoir systems with the magnitude rising with reduced prime volumes. The effect was markedly greater with the IC-HSVR. The IC-HSVR also transmitted larger bubbles, particularly with reduced prime volumes. There was no significant difference in microbubble transmission seen between the SC-SSVR-closed and SC-SSVR-open. The SC-SSVR transmits fewer microbubbles than the IC-HSVR during cardiotomy infusion and should be considered as the preferential system. Because both cardiotomy-reservoir systems transmitted microbubbles during cardiotomy infusion, particularly at the lower venous reservoir volume, it is important to use strategies to minimize cardiotomy microbubble infusion.
منابع مشابه
Massive air embolism in a Fontan patient.
Most institutions performing cardiopulmonary bypass for congenital heart disease patients use an integrated hard shell cardiotomy and venous reservoir attached to an oxygenator. It is of paramount importance that the integrated reservoir be vented so as not to cause pressurization. A pressurized sealed cardiotomy has been reported to occur secondary to issues with vacuum assisted venous drainag...
متن کاملVacuum-assisted venous drainage, angel or demon: PRO?
Vacuum-assisted venous drainage (VAVD) was proposed to optimize venous drainage during bypass through femoral venous cannulation. It is currently used in both adult and pediatric surgery when siphon gravity venous drainage is suboptimal. In pediatric surgery, the major advantages of VAVD are a significant decrease in cardiopulmonary bypass prime volume and an improved drainage with all collater...
متن کاملMicrobubble generation and transmission of medtronic's affinity hardshell venous reservoir and collapsible venous reservoir bag: an in-vitro comparison.
Microemboli are implicated in neurological injury; therefore, the extracorporeal circuit (ECC) should not generate microbubbles or transmit introduced air. The venous reservoir is the first component in the ECC designed to remove introduced air. The purpose of this study was to investigate the relative safety of two kinds of adult venous reservoirs--the closed soft-shell venous reservoir (SSVR ...
متن کاملGravity separation of pericardial fat in cardiotomy suction blood: an in vitro model.
Fat emboli generated during cardiac surgery have been shown to cause neurologic complications in patients postoperatively. Cardiotomy suction has been known to be a large generator of emboli. This study will examine the efficacy of a separation technique in which the cardiotomy suction blood is stored in a cardiotomy reservoir for various time intervals to allow spontaneous separation of fat fr...
متن کاملAn ultrasonic analysis of the comparative efficiency of various cardiotomy reservoirs and micropore blood filters.
The ability of 12 commercially available cardiotomy reservoirs to remove bubbles from aspirated blood was investigated by means of a simulated cardiopulmonary bypass circuit and an ultrasonic microbubble detector. Performance varied considerably. The number of gaseous microemboli remaining after passage of blood through the reservoir was reduced by (a) holding the blood in the reservoir, (b) re...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The journal of extra-corporeal technology
دوره 45 2 شماره
صفحات -
تاریخ انتشار 2013